Control Commands
optoNCDT ILR 1191
10.3.11 TD – Trigger Delay, Trigger Level
Format: TDx y [Enter]
Parameterizes behavior in external trigger mode (DF).
- x designates the delay in triggering a single measurement, in units of a millisecond.
- y designates the edge on arrival of which measurement will be triggered:
0 means that measurement is triggered on a falling edge (high-to-low transition)
1 means that measurement is triggered on a rising edge (low-to-high transition)
Request
TD
Set
TDx y
Parameter value
range x:
0 … 300.00 msec; resolution: 0.01 msec
Parameter value
range y:
0 oder 1
Standard
0.00 msec 0
10.3.12 SE – Error Mode
Format: SEx [Enter]
Parameterizes the behavior of the two switching outputs Q1 and Q2 and that of the analog output QA in the
event of failure to measure plus the state on completion of a single distance measurement.
Request
SE
Set
SEx
Value range
Parameter x Q1, Q2 (z=0)
Q1, Q2 (z=1)
QA
0
Latest value
Latest value
Latest value
1
High
Low
3 mA
2
Low
High
21 mA
Standard
1
The sensor performs no check for plausibility of a selected error mode.