Control Commands
optoNCDT ILR 1191
10.3.10 MF – Measurement Frequency [Hz]
Format: MFx [Enter]
Parameterizes the number x of measuring values to be emitted per second.
Request
MF
Set
MFx
Parameter value
range x:
1 … 2000; resolution: 1
Standard
2000
For example, MF1000 means that 1000 measuring values will be emitted each second. The time to measure
and, hence, the transmission of a measured result via the serial interface additionally depend on the setting
for parameter SA.
Data rate = Measurement frequency / Number of average value
Examples:
MF1000, SA1000: Data rate = 1 (1 measured value per second at the serial interface)
MF2000, SA1000: Data rate = 2 (2 measured values per second at the serial interface)
MF2000, SA20000: Data rate = 0.1 (1 measured value at the serial interface every 10 s))