Page 21
Distance Measurements
eddyNCDT 306x / 307x
Positioning the Target
Position the target within the sensor measuring
range.
The value for the start of the measuring range (SMR)
depends on the sensor. This value can be found in the
technical data of the sensor.
SMR
Sensor
20 mA
12 mA
4 mA
Target
Measuring range (MR)
SMR MMR EMR
Displace-
ment
10 V
5 V
0 V
Signal
SMR Start of measuring range
MMR Midrange
EMR End of measuring range
Distance Measurements
Switch to the
Measurement
menu.
Click the
Start measuring
button.
Statistic values are calculated in the web interface. Click-
ing onto the
start/stop measuring
button starts/
stops the calculation. At the beginning of a measure-
ment, the statistic values are reset. During a measure-
ment, the statistic values are updated with each new
data package received by the controller.