Page 16
Operation Using Ethernet
eddyNCDT 306x / 307x
3-Point Linearization
Menu
Settings
>
Characteristics/Linear-
ization
>
Carry out field linearization
.
If the sensor or the measurement object is changed by
the user, a calibration must be carried out before the
measurement. Here, use the following if possible:
- the original sensor mounting,
- the original measurement object.
i
Before a calibration is performed, the measuring
device should warm up for about 30 minutes.
Choose
3-point
for linearization and the desired
unit
.
Exemplary linearization using an ES-U3 sensor
Sensor balancing occurs via three distance points which
are specified by a comparison standard. You can freely
choose the linearization points within the sensor mea-
suring range.
Place the mea-
surement object
to the sensor in
point 1.
SMR
Sensor
Target
0 %
50 %
100 %
Measuring range
Enter the mea-
surement value
(1).
1
Confirm point 1 with
Submit
.
Repeat this procedure for the linearization points 2
and 3.
Click on the button
Linearize
. The system exe-
cutes the linearization.
You can permanently store the linearization result.
Select a memory location with
Select Field
characteristic
.
Enter a description for the linearization in the field
Set name
.
Click on the button
Save & activate
.