![MICRO-EPSILON eddyNCDT 3060 Quick Manual Download Page 19](http://html1.mh-extra.com/html/micro-epsilon/eddyncdt-3060/eddyncdt-3060_quick-manual_1785251019.webp)
Page 19
Operation Using Ethernet
eddyNCDT 306x / 307x
Limit Output
This function is available with controller DT3061, DT3071.
Menu
Settings
>
Outputs
>
Limit
value 1/2
.
The eddyNCDT 3061, 3071 can check the measurement result to
adjustable limits. This means that threshold values can be
monitored, impermissible tolerances detected and sorting criteria
realized.
The reference for the limit monitoring is selectable and applies to
the current characteristic.
Type: Relative | Peak-To-Peak | Dynamic.
Relative
The threshold values A/B refer to the set
Reference value
Peak-To-Peak
The threshold values A/B refer to the
peak-to-peak value calculated in blocks
(
Peak-to-Peak ∆t
parameter).
Dynamic
The threshold values A/B refer to a continuous-
ly calculated, moving average (
Average
∆t
).
Threshold A (A)
Threshold B (B)
Reference value
Relative
A
Peak-To-Peak
Dynamic
B
Peak-to-Peak
t
Average
value
A
B
A
B
Average
t
Peak-Peak value
References for limit monitoring