![MICRO-EPSILON eddyNCDT 3060 Quick Manual Download Page 17](http://html1.mh-extra.com/html/micro-epsilon/eddyncdt-3060/eddyncdt-3060_quick-manual_1785251017.webp)
Page 17
Operation Using Ethernet
eddyNCDT 306x / 307x
Here you can export all controller settings in a file or
reimport them from a file.
Controller settings Hardware filter, limit value outputs
Ethernet settings
IP address, subnet mask
Scaling Temperature Analog Output
This function is possible with the DT3061 and DT3071
controller.
Menu
Settings > Outputs > Temperature
.
The temperature output enables to output the controller
or sensor temperature.
Choose the types
Temperature sensor
or
Tem-
perature controller
.
Max. output range: 0 V ... 5 V
-20 °C
+180 °C
5 V
0 V
Output
1
Temp
0 °C
+70 °C
5 V
0 V
Output
2
Temp
Sensor temperature
Controller temperature
The accuracy of the temperature measurement depends
on the installation scenario. Reproducibility is high.
Select Characteristic
Menu
Settings
>
Characteristics/Linear-
ization
>
Current Characteristic.
The DT3060, DT3070 can save one field characteristic
curve. The DT3061, DT3071 can save up to four different
field characteristic curves, which are based on one a
factory calibration respectively.
Therefore, you can e.g.
store different target or
installation scenarios as in-
dividual characteristiccurve
and load them into the
controller for the desired
application. The
Type
field
informs you about the un-
derlying linearization type.
Via the menu Select characteristic, choose the
desired characteristic curve or linearization for your
measurement.
Import, Export
Menu
Settings >
System settings >
Manage settings.