R-Net LED and VR2
In case of an error the number of blinking light segments (1),
(2), (3) indicate the possible cause of the malfunction.
☞
Therefore observe table <
Error diagnostics
>.
☞
Note:
If the malfunction cannot be repaired and the vehicle
no longer operated, contact your national Meyra-Orto-
pedia sales partner.
Error diagnostics
LED
Cause of error
Fault correction
The battery voltage is too low.
Check the plug on the operating module
and the battery connections. Charge the
batteries.
The electrical connection to the left mo-
tor is open-circuit.
Check the motor connection cable, motor
cable plug and motor.
The electrical connection to the left mo-
tor is defective (short-circuit).
Check the motor connection cable, motor
cable plug and motor.
The electrical connection to the right
motor is open-circuit.
Check the motor connection cable, motor
cable plug and motor.
The electrical connection to the right
motor is defective (short-circuit).
Check the motor connection cable, motor
cable plug and motor.
The drive disable function is active.
The drive disable function is an electronic se-
curity function that prevents the wheelchair
from being driven when a battery charger is
connected.