MI 3242 MicroOhm 2A
Technical specification
39
10.2
Measurement parameters
Test current:
2
A
100
m
A
10
m
A
Max. power output:
4
W
0,2
W
20
mW
Output voltage:
9
V
DC
max.
Limits:
1
μΩ
… 199.9
Ω
Table 10.2: Measurement parameters
10.3
Voltage and frequency
10.3.1
Voltage
Measuring range (V)
Resolution (V)
Accuracy
0.0
49.9
0.1
50
550
1
(2 % of r 2 digits)
Result type........................................ True r.m.s. (trms)
Nominal frequency range.................. 0 Hz, 14 Hz
500 Hz
10.3.2
Frequency
Measuring range (Hz)
Resolution (Hz)
Accuracy
10.0
99.9
0.1
100
500
1
(0.2 % of r 1 digit)
Nominal voltage range...................... 10 V
550 V