MI 3242 MicroOhm 2A
Measurements
25
Figure 6.4: Examples of Resistance measurement results
Note:
Consider displayed warnings when starting the measurement!
6.2.1
Single Mode
Single mode makes a single bidirectional measurement. The instrument will measure
resistance in both directions (thermal EMF elimination). The main result displayed on
the display is an average (
2
R
R
R
).
Single measurement I/t plot
Figure 6.5: Single Mode
Note:
The Single Mode can be primarily used for measuring:
Relays
Switches
Connectors
Bus bars
Power distribution cable joints
Welding joints