MI 3242 MicroOhm 2A
Measurements
24
6.2 Resistance measurement
Test can be started from the Resistance window. Before carrying out a test the
parameters (Mode and Current) can be edited.
Figure 6.2: Resistance menu
Test parameters for Resistance measurement
Mode Sub-function
[Single,
Continuous, Auto, Inductive]
Current
Test current [2 A, 100 mA, 10 mA]
Keys:
TAB
Selects the field to be changed.
/
Modifies selected field.
TEST
Start the resistance measurement.
Test circuits for Resistance measurement
Figure 6.3 Resistance measurement connection
Resistance measurement procedure:
Select the RESISTANCE function.
Set the test parameters (mode and test current).
Set limits (optional). (See chapter
Connect device under test to the instrument (see figure 6.3).
Press the TEST key to start the measurement.
Press the TEST key to stop the measurement (not used in single mode).
Store the result by pressing MEM key (optional).