MI 3242 MicroOhm 2A
Technical specification
38
10 Technical specifications
10.1
Resistance measurement
Test current
Resistance range
Resolution
Accuracy
0.000 … 9.999
m
Ω
1
µ
Ω
10.00 … 99.99
m
Ω
10
µ
Ω
2
A
100.0 … 999.9
m
Ω
100
µ
Ω
0.00 … 99.99
m
Ω
10
µ
Ω
100.0 … 999.9
m
Ω
100
µ
Ω
1.000 … 9.999
Ω
1
m
Ω
100 mA
10.00 … 19.99
Ω
10
m
Ω
0.0 … 999.9
m
Ω
100
µ
Ω
1.000 … 9.999
Ω
1
m
Ω
10.00 … 99.99
Ω
10
m
Ω
10 mA
100.0 … 199.9
Ω
100
m
Ω
(0.25 % of r 2 digits)
Table 10.1: Resistance measurement ranges and accuracy
Test current accuracy ............... ±10% (smoothed DC)
Test duration............................. 1 s (single mode)
Test method.............................. 4-wire measurement
Note:
All data regarding accuracy is given for nominal (reference) environment
condition and forward and reverse measurements.
Inductive mode will introduce an undefined error if an EMF is present on the test
object.
The error in operating conditions could be at most the error for reference
conditions (specified in the manual for each function)
+
0.1% of measured
value + 1 digit, unless otherwise specified in the manual for particular function.
Noise rejection (50/60 Hz) on potential leads P1 - P2:
Test current
Max. noise level
Additional error
2 A
1.4 A
RMS
(R
x
< 500 m
Ω
)
100 mA
70 m
A
RMS
(R
x
< 10
Ω
)
10 mA
7 m
A
RMS
(R
x
< 100
Ω
)
≤
0.5 %