MI 3242 MicroOhm 2A
Settings menu
20
5.2 Initial settings
In this menu the following instrument parameters can be set to their initial values:
all measurement parameters;
LCD settings;
language;
Figure 5.3: Initial settings menu
Keys:
TAB
Selects YES or NO
TEST
Confirms selection.
The instrument will restart with the default settings (if YES is selected).
Returns to the Settings menu without changes (if NO is selected)
.
FUNCTION
SELECTOR
Returns to
Main menu
without changes.
5.3 Limits settings
With High and Low limit the user is allowed to set limit resistance values. Measured
resistance is compared against those limits. Result is validated only if it is within the
given limits.
Figure 5.4: Limits settings menu
Parameters for Limits settings:
Limit Hi
Limit Lo
Resistance limit value [OFF, 1.00
μΩ
- 10.00
μΩ
, 20.0
μΩ
- 100.0
μΩ
,
0.200 m
Ω
- 1.000 m
Ω
, 2.00 m
Ω
- 10.00 m
Ω
, 20.0 m
Ω
- 100.0 m
Ω
,
0.200
Ω
- 1.000
Ω
, 2.00
Ω
- 10.00
Ω
, 20.0
Ω
- 200.0
Ω
]
Keys:
UP/DOWN
Selects the High or Low limit to be modified.
TEST
Enters Limits modification screen.
FUNCTION
SELECTOR
Returns to
Main menu
without changes.
Figure 5.5: Limits modifications