Safety and operational considerations
7
Warnings related to measurement functions:
Resistance, voltage and frequency measurements
Resistance measurement should only be performed on de-energized objects!
Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!
When a resistance measurement has been performed on an inductive object,
automatic discharge may not be done immediately!
Do not connect test terminals to external voltage higher than 600 V (AC or DC) in
order not to damage the test instrument!
Notes related to measurement functions:
General
Warning
will appear on screen and the resistance test will not be performed if
voltages higher than 8 V (AC or DC) are detected between test terminals. There
will be no warning if all terminals are at the same potential.
PASS / FAIL indication is enabled when limit is set. Apply appropriate limit value
for evaluation of measurement results.
The internal memory is full!