![METREL MI 3100 S EurotestEASI Instruction Manual Download Page 73](http://html1.mh-extra.com/html/metrel/mi-3100-s-eurotesteasi/mi-3100-s-eurotesteasi_instruction-manual_1779038073.webp)
MI 3100 s, MI 3100 SE
Eurotest
EASI
Technical specifications
73
10.2 Continuity
10.2.1 Resistance R LOW
Measuring range according to EN 61557 is 0.16
1999
.
Measuring range R (
)
Resolution (
)
Accuracy
0.00
19.99
0.01
(3 % of r 3 digits)
20.0
199.9
0.1
(5 % of reading)
200
1999
1
Measuring range R+, R- (
)
Resolution (
)
Accuracy
0.0
199.9
0.1
(5 % of r 5 digits)
200
1999
1
Open-circuit voltage ................................ 6.5 VDC
9 VDC
Measuring current ................................... min. 200 mA into load resistance of 2
Test lead compensation .......................... up to 5
The number of possible tests .................. > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.
10.2.2 Resistance CONTINUITY
Measuring range (
)
Resolution (
)
Accuracy
0.0
19.9
0.1
(5 % of r 3 digits)
20
1999
1
Open-circuit voltage ................................ 6.5 VDC
9 VDC
Short-circuit current ................................ max. 8.5 mA
Test lead compensation .......................... up to 5