MI 3100 s, MI 3100 SE
Eurotest
EASI
Measurements
41
Maximum test current is I
(trip-
out current) or end value in case the RCD didn’t trip-out.
Trip-out current measurement procedure
Select the
RCD
function using the function selector switch.
Set sub-function to
RCD I
using
UP / DOWN
keys.
Set test
parameters
(if necessary).
Connect
test cable to the instrument.
Connect
test leads to the item to be tested (see
Press the
TEST
key to perform the measurement.
Store
the result by pressing the
MEM
key (MI 3100 SE only).
Figure 5.20: Trip-out current measurement result example
Displayed results:
I
............ trip-out current
Uci
....... contact voltage at trip-out current I or end va
lue in case the RCD didn’t trip
t
............ trip-out time