MD 9273 Leakage Clamp
Technical Specifications
23
4
Technical Specifications
4.1
General specifications
–
Environment conditions
Measurement category clamp-meter
CAT III
600V
Measurement category test lead
CAT III 600V
Maximum current
2
100 A
Pollution degree
2
Operation
Indoor
Altitude
< 2000 m
Operation temperature
0 ºC –
40 ºC (<80 % H.R., <10 ºC
without
condensation)
Storage temperature
10 ºC –
60 ºC (<70 % H.R., extract batteries)
Temperature coefficient:
0.1 (of s
pecified precision) / ºC (<18 ºC or >28 º
C)
Functioning principle:
Sampling the signal at high speed
Data update on screen:
Twice a second
Screen:
3 3/4 + 3 digits LCD screen
Range selection:
Automatic and manual selection
Over-range indication:
-0L- is shown on the LCD screen, fast beep
Polarity indication:
“
-
“ is shown automatically
E.M.C.:
Emmission: Class B
Immunity: Portable test and measurement
equipment
Meets EN 61326-1:2013, EN 61326-2-2:2013
Safety:
Meets EN 61010-1:2010, EN 61010-2-32:2012,
EN 61010-2-033:2012, EN 61010-031:2015
Bluetooth®:
v3.0
Power supply:
3 V
Type of batteries:
2 x 1,5 V AA alkaline batteries
Consumption:
5 mA average, depending on function.
50 mA average i
n Bluetooth® mode
Duration of the batteries:
200 - 400 h (batteries 2000 mAh)
20
–
40 h
(Bluetooth® mode)
Dimensions:
185 mm x 62 mm x 42 mm
Weight:
Approx. 210 g (without battery)
4.2
Measurement specifications
Current and voltage support automatic ranging within the same function.
2
This equipment cannot be exposed to currents above the limit