MD 9273 Leakage Clamp
Technical Specifications
26
Peak Value
Function
Sampling
time
Range
Resolution
Accuracy
mA,
A
25 µs
40 mA
0.01 mA
±
5
% of reading ±
5 digits
400 mA
0.1 mA
±
5
% of reading ±
5 digits
4 A
0.001 A
±
5
% of reading ±
5 digits
40 A
0.01 A
±
5
% of reading ±
5 digits
80 A
0.1 A
±
5
% of reading ±
5 digits
100 A
0.1 A
±
5 % of reading
± 5 digits
V
40 V
0.01 V
±
5
% of reading ±
5 digits
400 V
0.1 V
±
5
% of reading ±
5 digits
600 V
1 V
±
5
% of reading ±
5 digits
Inrush
Function
RMS cycle time
calculation
Range
Resolution
Accuracy
mA,
A
10ms (Half cycle)
40 mA
0.01 mA
± 1.5 % of reading ± 3 digits
400 mA
0.1 mA
± 1.5 % of reading ± 3 digits
4 A
0.001 A
± 1.5 % of reading ± 3 digits
40 A
0.01 A
± 1.5 % of reading ± 3 digits
80 A
0.1 A
± 2 % of reading ± 3 digits
100 A
0.1 A
± 5 % of reading ± 3 digits
V
40 V
1 V
± 1 % of reading ± 1 digits
400 V
1 V
± 1 % of reading ± 1 digits
600 V
1 V
± 1 % of reading ± 1 digits
VFD
Function
Range
Resolution
Accuracy
V
40 V
0.01 V
± 2
% of reading ± 4 digits
400 V
0.1 V
± 2
% of reading ± 2 digits
600 V
1 V
± 2
% of reading ± 2 digits
Note: Specification is valid
for PWM signals in the range of 1 kHz - 5 kHz and fundamental frequency in the
range of 20 - 100 Hz.