MD 9273 Leakage Clamp
Description of Test Functions
22
When recording the value, the difference between the peak value and the maximum value should be taken
into account: the maximum value holds the highest measured value on the screen, whilst the peak value
shows the highest absolute value of the waveform.
MIN Mode
In this mode, the minimum measured value is shown on the screen. The
symbol will be shown. Before
using this function, the clamp should be connected properly, see chapter Description of Test Functions
for
details.
3.5
Bluetooth
®
Mode Functions
Activate the mode by setting the rotary switch to the position. In this mode, the MD 9273 is controlled by
Bluetooth® connected device,
a
Metrel® i
nstrument or other application.
After switching to Bluetooth mode, the screen will show
“bt Ini” and
few seconds later “bt”. This means the
MD 9273 is ready to be paired with control device. Follow the user manual of control device. When paired,
the symbol appears on the screen.
The paired device now controls the MD 9273. If a measuring function is set,
message “ME
A
” appears on the
screen. If a recording
function is set, message “rEC” is presented
.
Follow the user manual of the instrument to which the MD9273 is linked for details on supported test
functions and their settings.