General description
8
1.1.2 Markings on the instrument
Read the Instruction manual with special care to safety
operation«. The symbol requires an action!
Mark on your equipment certifies that it meets European Union
requirements for EMC, LVD, and ROHS regulations.
This equipment should be recycled as electronic waste.
1.1.3 Warnings related to safety of batteries
When connected to an installation, the instruments battery compartment can contain
hazardous voltage inside! When replacing battery cells or before opening the battery/fuse
compartment cover, disconnect any measuring accessory connected to the instrument
and turn off the instrument,
Ensure that the battery cells are inserted correctly otherwise the instrument will not
operate and the batteries could be discharged.
Do not recharge alkaline battery cells!
Use only power supply adapter delivered from the manufacturer or distributor of the test
equipment!
1.1.4 Warnings related to safety of measurement functions
Insulation resistance
Insulation resistance measurement should only be performed on de-energized objects!
Do not touch the test object during the measurement or before it is fully discharged! Risk
of electric shock!
Continuity functions
Continuity measurements should only be performed on de-energized objects!