MI 3152(H)
Eurotest
XC (2,5 kV)
General description
12
Connect test cable to the instrument.
Connect test leads to the object under test, see
Figure 1.1
and
Figure 1.2
.
Touch
test probe for at least 2 seconds.
If PE terminal is connected to phase voltage the warning message is displayed,
instrument buzzer is activated and further measurements are disabled in Z loop, Zs rcd,
RCD tests and Auto test sequences.
Notes
PE test terminal is active in the RCD tests, Z loop, Zs rcd, Z line, dU and Voltage
measurements and Auto test sequences only!
For correct testing of PE terminal, the
key has to be touched for at least 2 seconds.
Make sure to stand on non-isolated floor while carrying out the test, otherwise test result
may be wrong!