7.16
Ro – Specific earth resistance
Figure 7.55: Earth Ro menu
Measurement parameters / limits
Distance
Distance between probes
[0.1 m ... 30.0 m] or [1 ft ... 100
ft]
Connection diagram
Figure 7.56: Specific earth resistance measurement
Measurement procedure
Enter the
Ro
function.
Set test parameters / limits.
Connect A 1199 adapter to the instrument.
Connect test leads to earth probes, see
Figure 7.56
.
Start the measurement.
Save results (optional).