BAM 1020-9800 Manual Rev W.docx
Page 95
If the error status is included in the custom Query array (QC), it will appear as a decimal error
code as shown below. Each decimal code value corresponds to one of the twelve regular
BAM 1020 error or alarm types as described in Section 7.2.
Decimal Code
Error Flag
Description
0
none
No error
1
T
Tape System Errors
2
C
Beta Count Error
4
D
Deviant Membrane Density Alarm
8
P
Pressure Drop Alarm
16
F
Flow Errors
32
N
Nozzle Error
64
R
Reference Error, Membrane Timeout
128
L
Power Failure
256
I
Internal Error, Coarse Link Down
512
M
Maintenance Flag
1024
U
Telemetry Fault
2048
E
External Reset Error
A data integrity checksum is included at the end of each Query array, after the delimiter
asterisk
character ‘*’. The checksum is the 16-bit arithmetic sum of all characters in the line
up to but not including the asterisk.
Note:
Digital data loggers may be programmed to use the <STX> (hex 02) character instead
of the <ESC>, to prevent the echoing of the command back to the logger.