BAM 1020-9800 Manual Rev W.docx
Page 106
BAM 1020 Audit Sheet
Model:
BAM 1020
Serial Number:
Audit Date:
Audited By:
Flow Audits
Flow Reference Standard Used:
Model:
Serial
No:
Calibration
Date:
Temperature Standard Used:
Model:
Serial
No:
Calibration
Date:
Barometric Pressure Standard
Used:
Model:
Serial
No:
Calibration
Date:
Leak Check Value:
as
found:
lpm
as left:
lpm
BAM
Ref. Std.
BAM
Ref. Std.
Ambient Temperature:
as
found:
C
C
as left:
C
C
N/A
Barometric Pressure:
as
found:
mmHg
mmHg
as left:
mmHg
mmHg
Flow Rate (Actual
Volumetric):
as
found:
lpm
lpm
as left:
lpm
lpm
N/A
Flow Rate (EPA Standard):
as
found:
slpm
slpm
as left:
slpm
slpm
N/A
Mechanical Audits
Pump muffler unclogged:
as found
as left
PM10 particle trap clean:
as
found
as left
N/A
Sample nozzle clean:
as found
as left
PM10 drip jar empty:
as
found
as left
N/A
Tape support vane clean:
as found
as left
PM10 bug screen clear:
as
found
as left
N/A
Capstan shaft clean:
as found
as left
PM2.5 particle trap clean:
as
found
as left
N/A
Rubber pinch rollers clean:
as found
as left
Inlet tube water-tight seal OK:
as
found
as left
Chassis ground wire
installed:
as found
as left
Inlet tube perpendicular to BAM:
as
found
as left
Analog Voltage Output Audit
N/A
DAC Test
Screen
BAM Voltage
Output
Logger Voltage
Input
0.000 Volts
Volts
Volts
0.500 Volts
Volts
Volts
1.000 Volts
Volts
Volts
Membrane Audit
Flow Control Range
LAST m (mg):
Flow Set
point
BAM Flow
ABS (mg):
15.0 LPM
Difference
(mg):
16.70 LPM
% Difference:
18.4 LPM