BAM 1020-9800 Manual Rev W.docx
Page 50
Datalog RH:
If YES is selected, the filter RH values will be logged on channel 4 of the
BAM 1020. Select YES if you do not have any external sensors attached to channel 4.
Properly Regulated Channel 4 Hourly Filter RH Graph
Delta-T Control:
If Delta-T Control is enabled, the BAM 1020 will continuously measure
the difference between ambient temperature (requiring either BX-592 or BX-596) and the
filter temperature. If enabled, the Smart Heater will not apply excess heat if the difference
between ambient temperature and the filter temperature exceeds the set point even if the
filter RH exceeds its set value. Delta-T Control is not enabled for the US-EPA PM
2.5
, PM
10
, or
PM
10-2.5
configurations of the BAM 1020. It is not enabled for the EU configurations of the
BAM 1020.
Delta-T Set point:
Only used if Delta-T Control is enabled. This can be set from 1 to 99
degrees C. If the Delta-T exceeds this set point, the Smart Heater will turn down to low power
mode, regardless of the RH level. An
“N” alarm is logged any time this happens. Note: There
is often a few degrees of Delta-T measured even if the heater is OFF, due to mild heating
effect of the BAM 1020 itself. Frequent errors will be logged if the set point is too low. Set the
value to 99 if Delta-T control is not used.
Datalog Delta-T:
If YES is selected, the Delta-T values will be logged on met channel 5 of
the BAM 1020, as long as no other external sensors are attached to channel 5. Note:
The
measured Delta-T values may still be logged even if Delta-T control is set to NO
. Many BAM
1020 users log the Delta-T for informational purposes.
6.10
QUERY Setup Screen
– Custom Data Array Setup
The SETUP > QUERY screen allows the user to configure a custom digital data array for the
new Query Output, or for the European BH Bayern-Hessen protocol. The user can select
exactly which data parameters appear in the array, and in what exact order they appear. The
custom array can be as simple as a single concentration value, or it can be very
comprehensive including some parameters that are not even available in any of the standard
data files, such as the hourly span membrane check value. The setup of this custom array
does not affect any of the standard BAM 1020 data arrays. The query output is available in
firmware revisions 3.6.3 and later and requires the Report Processor. The BH protocol is
available in revision 5 series European firmware. A separate technical document for the
Bayern-Hessen protocol functions is also available. See Section 9.9 for instructions about
how to retrieve the Query output files.