BAM 1020-9800 Manual Rev W.docx
Page 54
7
MAINTENANCE, DIAGNOSTICS and TROUBLESHOOTING
This section provides information about routine maintenance, identifying errors and alarms,
and performing diagnostic tests on the BAM 1020. The TEST menu functions are also
described in this section.
Met One Instruments also publishes a comprehensive array of technical bulletins that covers
additional information about subsystem troubleshooting, upgrades, and repairs. These are
available in the “BAM Users” section of our website, or by e-mail request from the Technical
Service department.
7.1 Met One Recommended Periodic Maintenance Table
The following table shows the recommended interval for the regular BAM 1020 maintenance,
field check, and service tasks. Special tools are not required for any of the routine BAM
service tasks on less than yearly intervals. Met One recommends the BX-308 and BX-344
kits for non-routine service and repairs such as nozzle removal and detector tests. Complete
instructions are included.
Maintenance Item
Period
Nozzle and vane cleaning.
Monthly
Leak check.
Monthly
Flow system check/audit.
Monthly
Clean capstan shaft and pinch roller tires.
Monthly
Clean PM10 inlet particle trap and PM2.5 cyclone particle trap.
Monthly
Download and save digital data log and error log.
Monthly
Compare BAM 1020 digital data to external analog data logger data, if used.
Monthly
Check or set BAM real-time clock.
Monthly
Replace filter tape roll.
2 Months
Run the SELF-TEST function in the TAPE menu.
2 Months
Download and verify BAM 1020 settings file.
Quarterly
Complete flow system calibration.
Quarterly
Completely disassemble and clean PM10 inlet and PM2.5 cyclone.
Quarterly
Replace or clean pump muffler.
6 months
Test filter RH and filter temperature sensors.
6 months
Test smart heater function.
6 months
Clean internal debris filter.
12 Months
Remove and check membrane span foil.
12 Months
Beta detector count rate and dark count test.
12 Months
Clean vertical inlet tube (BX-344 cleaning kit).
12 months
Test analog DAC output, if used.
12 Months
Replace lithium battery if necessary.
12 Months
Rebuild vacuum pump.
24 months
Replace nozzle O-ring.
24 months
Replace pump tubing, if necessary.
24 Months
Factory recalibration is not required except for units sent for major repairs.
---