BAM 1020-9800 Manual Rev W.docx
Page 73
menu and allow the sensor to equilibrate for at least five minutes, then compare the BAM RH
reading on the display to your reference RH device. To calibrate the sensor, enter the
reference value into the BAM display and press CAL to change the BAM value to match.
The RESET key can be used to remove all previous field calibrations from the sensor
and restore the default factory calibration.
Do not press the CAL key after RESET, or
whatever value happens to be in the REFERENCE field will be calibrated.