BAM 1020-9800 Manual Rev W.docx
Page 37
will not change to match the STD until after you have entered all three flow calibration
points, since it is done on a slope.
6. Press the NEXT key to select the second flow point of 18.4 L/min. Allow the flow to
stabilize again, then enter the value from your standard device and press the CAL key.
Note: If the BAM 1020 is unable to achieve flow regulation at the 18.4 L/min point, this
could be an indication that the vacuum pump needs to be serviced.
7. Press the NEXT key to select the third flow point of 16.70 L/min. Allow the flow to
stabilize again, then enter the value from your standard device and press the CAL key.
8. After this third flow point is calibrated, the BAM flow reading will change to show the
corrected flow, then the BAM 1020 will quickly re-regulate the flow to 16.70 L/min
based on the new calibration. The BAM 1020 flow reading should now match your flow
standard device at 16.70 ± 0.1 L/min. Exit the calibration menu.
Resetting Flow Calibrations:
If the BAM 1020 flow, temperature, or pressure readings do not correctly change to match
your standard device during the above calibration process, or if multiple calibrations are
required in order to get a good match, then the BAM flow calibrations may need to be reset.
This case has sometimes been observed during the first flow calibration after a BAM 1020
firmware update.
Select a parameter and press the DEFAULT soft key to clear out all previous calibration
factors from that parameter and replace them with the original factory calibration factor.
Default all of the AT, BP, and flow parameters, then try again to calibrate them to your
standards. You may also need to reset the filter RH and filter temperature sensor calibrations.
The default factory calibration factors should be very close to the correct values.
Actual flow calibrations in units with older firmware:
BAM 1020 units with firmware revisions 2.58 and earlier had a different TEST > FLOW
screen, as shown below. These units are calibrated as described above, except that the flow
calibration is performed at only a single point of 16.70 L/min. The NEXT key selects the AT or
BP parameter to be calibrated, and the PUMP ON key selects the flow point for calibration.
The ADJUST/SAVE key calibrates the selected parameter to the reference value.
Old Format of the Actual Flow Calibration Screen
ACTUAL FLOW CALIBRATION MODE
F1= RESTORE DEFAULT
BAM REFERENCE
AMBIENT TEMPERATURE: 23.8 C 23.4 C
BAROMETRIC PRESSURE: 741 mmHg 742 mmHg
VOLUMETRIC FLOWRATE: 16.7 lpm 16.9 lpm
ADJUST/SAVE NEXT PUMP ON EXIT