Calibration Menu
You may run any of the calibration procedures when the feeder is the Idle state. The
hopper should typically be empty. The only two meaningful procedures with material in
the hopper are Weight Procedure and Material Procedure. If the
Allow Calibr
logical
input is Off you will not be able to accept any results.
Zeroing Procedure
The purpose of the
Zeroing Procedure
is to have zero weight indication when the
hopper is empty. The
Zero Weight
parameter is updated. The hopper must be empty,
and preferably cleaned when you perform this procedure.
When you push
Start
, The controller will take the Stable Weight as the new
Zero Weight
.
If you do not wish to run the Zeroing Procedure, push the
Cancel
button to return to the
Calibration Menu. However, if you are ready to begin the
Zeroing Procedure,
push
Start
.
The display will show the elapsed time of the test and an indication on the stability of the
scale.
When the procedure is complete the screen will show the current and new
Zero Weight
values, as well as the change in the
Zero Weight
value as a percent of
Design Weight
.
If you wish to accept the new
Zero Weight
value, push
Accept
. Otherwise, push
Cancel
.
Pushing either button will return you to the calibration menu.
35.00.EX O&M Manual
47