31
Use the
Del
soft key to delete the entire calibration record.
Note: This Del key only removes the record from
viewing by the MFT; it does not clear memory space. Connect the MFT to Meriam DMS or PC utility (pending)
and use the clear MFT calibration data function to clear memory.
Use the
Done
soft key to end the Viewing session
Use the
Next
soft key to move to the next saved point for the active tag number
Use the
Back
soft key to go back to the prior screen
Clear Cal Memory:
To clear all calibration procedure / result files,
Select
the
Clear Cal Memory
menu option
under the MFTs
Calibrate
key (see the Calibrate Key section of this manual for more details). A delete
confirmation screen is provided to prevent unintended clearing of all documentation files.
Cal
Results: PT-307
Single variable
As found
Point 1 of 6
In: 0.090 inW20C
Out: 3.999 mA
Error: -0.096 %
Pass
Del Done Next Back
View Calibration Results
1.888.475.5235