103
4.0 Specifications
Input impedance: > 10 megohm
Warm-up time:
≤
30 Seconds to full rated accuracy
Ohm Measurement Accuracy:
Temperature Effect:
≤
0.001 ohm /
°
C
RTD Measurement Accuracy: See RIO Accuracies Table
RTD Simulation Accuracy: See RIO Accuracies Table
Resolution:
±
0.1
°
or
±
1
°
for RTDs
Resistance range: 0 – 4000 ohms
Temperature:
Storage: -40
°
F to 140
°
F (-40
°
C to 60
°
C)
Operating: 14
°
F to 122
°
F (-10
°
C to 50
°
C)
Connectors – Measure temperature: standard connector for reference RTD or Measure Cable for general purpose
measurements on installed RTDs
Simulate resistance: User Cable with connector and two alligator clips
Power: supplied from MFT (simulate/source mode only)
Units:
°
F,
°
C,
°
R,
°
K, ohms
Weight: 3 ozs
5.0 Accessories for RIO4000
P/N A900028-90502
Simulation Cable for RTD 2-Wire simulation.
P/N A900028-90200
Mating plug connector for use with customer’s dedicated RTD
P/N A900028-90500
Measure cable assembly; RTD connector, 5 ft. copper cable, four (4) alligator clips
P/N A900028-90201
Pt 100 RTD, DIN Class "A", 0.00385 alpha, 4 leads, 1/4"
O.D., SS probe, 10" probe length, 3.5" handle and 5 ft coiled cable with connector for
RIO4000 temperature module
1.888.475.5235