dip the probe in a water bath and check if the temperature reading differs unacceptably from a
laboratory approved reference thermometer.
The thermometer should then be submitted to the manufacturer or retailer for recalibration.
The above recommendation does not replace any legal requirements. Legislation, rules and
regulations, and guidelines for the use, functionality and accuracy of the fever thermometer should
always be complied with.
TROUBLESHOOTING
Error message
Problem
Action
Measured temperature is
less than 32.0°C (90.0°F).
Switch off the thermometer. Wait one minute, start
the thermometer and attempt another measure-
ment with good surface contact.
Measured temperature ex-
ceeds 42.9°C (109.9°F).
Switch off the thermometer. Wait one minute, start
the thermometer and attempt another measure-
ment with good surface contact.
The product is not working
correctly.
Remove the battery, wait one minute, replace the
battery, and try to use the thermometer again. Con-
tact the retailer if the fault persists.
Low battery charge.
Replace battery.
ELECTROMAGNETIC COMPATIBILITY
The product complies with the EMC requirements in accordance with the international standard
IEC 60601-1-2 if the specified conditions in the table below are met. The product is an electrical
medical product and as such is subject to special requirements concerning electromagnetic
compatibility, which requirements must be indicated in the instructions for the product. Portable
and mobile radio communication equipment can affect the product. Do not use the product
together with unapproved accessories – this can have a negative effect on the product and change
its electromagnetic compatibility. Do not use the product in the immediate vicinity of electrical
equipment or between different electrical equipment.
GUIDELINES & MANUFACTURER’S CERTIFICATE CONCERNING
ELECTROMAGNETIC RADIATION
The product is intended to be used in electromagnetic environments in accordance with the
specifications below. It is incumbent on the customer or the person using the product to
ensure that the electromagnetic environment in which the product is used complies with these
specifications.