background image

 
 
 
 
 

Using the  

 

ISOTHERM 

 

2000 

 
 
 
 
 
 
 
 
 
 

    

   

 

              W. L. Walker Co., Inc. 

      Tulsa,Oklahoma

 

 

(888) 397- 9255 

 

Summary of Contents for ISOTHERM 2000

Page 1: ...Using the ISOTHERM 2000 W L Walker Co Inc Tulsa Oklahoma 888 397 9255...

Page 2: ...s DEPRESSED The display will show 8 8 8 8 8 8 8 This demonstrates that all of the display segments are still working After the initial display the ISOTHERM 2000 will switch to displaying the current t...

Page 3: ...SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY AVERTISSEMENT LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA S CURIT INTRINS QUE The probe and cable assembly on the ISOTHERM 2000 is designe...

Page 4: ...se the patch cable that is provided with the calibration tool Using the Calibration Tool 2000 While in the calibration mode you can adjust the temperature up and down with the Calibration Tool 2000 Pr...

Page 5: ...ture If you don t wait 5 minutes or more you won t get a good calibration Adjust the ISOTHERM s temperature until it matches the bath temperature See Using The Calibration Tool 2000 above to find out...

Page 6: ...nt to replace your battery immediately to prevent down time for the unit The OPEn display could appear in the following cases 1 There is no Probe Cable assembly attached to the ISOTHERM 2 The Probe Ca...

Page 7: ...o the ISOTHERM the ISOTHERM plugged into the ISOTHERM 2000 2000 Using incorrect procedure Review the ISOTHERM MARK calibration instructions I can t get a good Temperature hasn t stabilized Wait about...

Page 8: ...________________________________________________________________ ________________________________________________________________ WALKER LAB USE Date Tulsa ____ ____ ____ by __________ Description of...