Operating the ISOTHERM 2000

 

 

WARNING: 

Static Electricity can cause an EXPLOSION! 

BEFORE opening a tank hatch, attach the grounding clamp to the hatch cover. 

After you are done, close the tank hatch, THEN remove the grounding clamp. 

 

WARNING: 

REPLACE BATTERY WITH ANY GENERAL PURPOSE OR HEAVY DUTY 9-VOLT CELL.  FOR 

LONGER LIFE USE DURACELL 1604, ENERGIZER NO. 522, RADIOSHACK NO. 23-653, 

EVEREADY NO. 1222, RAYOVAC NO. A1604. 

 
Brief Description

 

The ISOTHERM 2000 is a hand-held portable digital thermometer designed specifically for use in 
hazardous locations.  The ISOTHERM 2000 is built to be durable and easy to operate.  It displays 
temperature readings on a 7-digit LCD with a backlight for easy reading in all viewing situations. 
 
Turning the ISOTHERM 2000 on 

PRESS AND HOLD 

the power button to keep the Isotherm power on.  

When the 

POWER

 button is DEPRESSED. The display will show: 

8.8.8.8.8.8.8 

This demonstrates that all of the display segments are still working.  After the initial display, the 
ISOTHERM 2000 will switch to displaying the current temperature (as measured by the tip of the 
attached probe). 
 
ISOTHERM 2000 Instructions 
The ISOTHERM 2000 can read temperature in either Fahrenheit or Celsius.  It also features a 
digital calibration tool that allows precision calibration without "tweaking pots" or doing clumsy 
calibration calculations. 
 
ISOTHERM 2000 KEYPAD 

 

Taking a temperature reading 

 

WARNING: 

Static Electricity can cause an EXPLOSION! 

BEFORE opening a tank hatch, attach the grounding clamp to the hatch cover. 

After you are done, close the tank hatch, THEN remove the grounding clamp. 

 
Take the stainless steel probe and place it in the fluid to be measured.  After the probe has been 
in the fluid for approximately 30 to 45 seconds, turn on the ISOTHERM 2000 and read the current 
temperature.  For the greatest accuracy, leave the probe in the fluid for five minutes or more. 
 
 
 
 
 
 
 
 

Summary of Contents for ISOTHERM 2000

Page 1: ...Using the ISOTHERM 2000 W L Walker Co Inc Tulsa Oklahoma 888 397 9255...

Page 2: ...s DEPRESSED The display will show 8 8 8 8 8 8 8 This demonstrates that all of the display segments are still working After the initial display the ISOTHERM 2000 will switch to displaying the current t...

Page 3: ...SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY AVERTISSEMENT LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA S CURIT INTRINS QUE The probe and cable assembly on the ISOTHERM 2000 is designe...

Page 4: ...se the patch cable that is provided with the calibration tool Using the Calibration Tool 2000 While in the calibration mode you can adjust the temperature up and down with the Calibration Tool 2000 Pr...

Page 5: ...ture If you don t wait 5 minutes or more you won t get a good calibration Adjust the ISOTHERM s temperature until it matches the bath temperature See Using The Calibration Tool 2000 above to find out...

Page 6: ...nt to replace your battery immediately to prevent down time for the unit The OPEn display could appear in the following cases 1 There is no Probe Cable assembly attached to the ISOTHERM 2 The Probe Ca...

Page 7: ...o the ISOTHERM the ISOTHERM plugged into the ISOTHERM 2000 2000 Using incorrect procedure Review the ISOTHERM MARK calibration instructions I can t get a good Temperature hasn t stabilized Wait about...

Page 8: ...________________________________________________________________ ________________________________________________________________ WALKER LAB USE Date Tulsa ____ ____ ____ by __________ Description of...

Reviews: