EN
74
GUIDELINES & MANUFACTURER’S CERTIFICATE CONCERNING
ELECTROMAGNETIC IMMUNITY
The product is intended to be used in electromagnetic environments in accordance with the
specifications below.
It is incumbent on the customer or the person using the product to ensure that the electromagnetic
environment in which the product is used complies with these specifications.
Immunity test
Test level as per IEC
60601
Compliance level
Guidelines for electromagnetic
environment
Line carried ra-
dio frequency
interference IEC
61000-4-6
3 Vrms 150 kHz to
80 MHz
Not applicable
Portable and mobile radio commu-
nication equipment must not used
nearer to any part of the product
or its leads than the recommended
minimum distance in accordance
with the equation applicable for
the transmitter in question.
Recommended minimum distance
where P is the maximum output
specified by the transmitter manu-
facturer in watts (W) and d is the
recommended minimum distance
in metres (m).
The field strength from a perma-
nent radio transmitter, determined
by electromagnetic measurement
in situ, should be less than the spe-
cified compliance level for the re-
spective frequency range.
Interference can occur in the vi-
cinity of equipment marked with
the symbol:
Emitted radio
frequency in-
terference IEC
61000-4-3
10 V/m 80 MHz to
2.7 GHz
10 V/m
Radio commu-
nication equip-
ment
IEC 61000-4-3
380 MHz, 27 V/m
450 MHz, 28 V/m
710 MHz,
745 MHz,
780 MHz 9 V/m
810 MHz,
870 MHz,
930 MHz 28 V/m
1720 MHz,
1845 MHz,
1970 MHz 28 V/m
2450 MHz, 28 V/m
5240 MHz,
5500 MHz,
5785 MHz 9 V/m
380 MHz, 27 V/m
450 MHz, 28 V/m
710 MHz,
745 MHz,
780 MHz 9 V/m
810 MHz,
870 MHz,
930 MHz 28 V/m
1720 MHz,
1845 MHz,
1970 MHz 28 V/m
2450 MHz, 28 V/m
5240 MHz,
5500 MHz,
5785 MHz 9 V/m
80 MHz to 800 MHz
800 MHz to 2.7 GHz
Summary of Contents for 013697