Menu mode
6. Accept the selection by pressing the
confirmation key. The adjustment op-
tions are highlighted.
Time
12.05.2012
12:00
GMT 00:00
Date
Time zone
Daylight savings
7. Set daylight savings to off (
) or on
(
) with the turn control – in this case
on (
). Save the setting by pressing
the confirmation key.
Time
12.05.2012
12:00
GMT 00:00
Date
Time zone
Daylight savings
Daylight saving time and standard time are not changed automatically. For this rea-
son, please keep in mind to adjust them at the beginning of each period.
8. Now, set date (day, month year) and
time (hours, minutes) in the same way.
Accept the selection by pressing the
confirmation key.
Time
Time zone
Daylight savings
.
.
2012
12:00
GMT 00:00
Date
27
05
7.5
Calibration
The appliances are calibrated and adjusted at the factory. In case readjustment should
be necessary later on – for example due to influence of the chamber load – temperature
and pressure can be adjusted according to customer requirements with three individually
selected adjustment points:
►
Cal1Adjustment at low value
►
Cal1Adjustment at middle value
►
Cal1Adjustment at high value
To guarantee perfect control, we recommend to calibrate the appliance once a year.
7.5.1 Temperature calibration
For temperature adjustment, you will need a calibrated reference measuring device.
Each thermoshelf can be calibrated individually. This is only possible with the
AtmoCONTROL software (see AtmoCONTROL manual for description).