40
D39374 | Date 08/2018
Operation and control
2. Accept the selection by pressing the
confirmation key. The
min
setting (un-
dertemperature protection) is automati-
cally activated.
ALARM
0.0
K
min
°C
auto
max
120
.0°C
100
0
.
+-
3. By turning the turn control, adjust the
desired lower alarm limit value, in the
example on the right 160 °C.
If no undertemperature protection limit
is required, set the lowest temperature.
ALARM
0.0
K
min
°C
auto
max
120
.0°C
160
0
.
+-
4. Press the confirmation key to confirm.
The
max
display (overtemperature
protection) is activated.
ALARM
0.0
K
min
°C
auto
max
°C
160
0
.
120
0
.
+-
5. By turning the turn control, adjust the
desired upper alarm limit value, in the
example on the right 190 °C.
The monitoring temperature must be
set sufficiently high above the maximum
set temperature. We recommend 5 to
10 K.
ALARM
0.0
K
min
°C
auto
max
°C
160
0
.
190
0
.
+-
6. Accept the upper alarm limit value
by pressing the confirmation key. The
setting of the automatic temperature
monitor (ASF) is automatically activated
(
auto
).
ALARM
max
190
.0°C
0.0
K
min
160
.0°C
auto
+-
7. With the turn control, select ON (
) or
OFF (
).
ALARM
max
190
.0°C
0.0
K
min
160
.0°C
auto
+-
8. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
ALARM
K
min
°C
auto
max
190
.
0°C
160
0
.
0.3
+-
9. With the turn control, adjust the desired
tolerance band, e.g. 5.0 K.
We recommend a tolerance band of 5
to 10 K.
ALARM
K
min
°C
auto
max
190
.
0°C
160
0
.
5.0
+-