46
D39374 | Date 08/2018
Malfunctions, warning and error messages
Error description
Cause of errors
Rectifying errors
See
Error message
T:E-3
in the temperature
display
TEMP
T:E-3
Set 37.0
°C
37
.4
°C
Temperature operating
sensor is defective.
►
Switch off appli-
ance
►
Contact customer
service
Page 2
Error message
AI E-3
in the temperature
display
TEMP
AI E-3
Set 37.0
°C
37
.4
°C
Temperature monitoring
sensor is defective.
►
Switch off appli-
ance
►
Contact customer
service
Page 2
Error message
E-3
in the temperature
display
TEMP
Set 45.0
°C
E-3
°C
Sensor defective
►
Switch off appli-
ance
►
Remove the cham-
ber load
►
Contact customer
service
Page 2
Error message
E-7
in
the pressure display
VACUUM
mb
Set 800
mb
In 1
E-7
Pressure sensor defective.
No pressure control pos-
sible.
►
Contact customer
service
Page 2
Summary of Contents for VO Series