background image

 

Operation Manual 

MELA

quick

®

12

 

Page 18 

7.8 

Sliding door cannot be unlocked ___________________________________________

 

Pressure operated lock of sliding door (1) 

As soon as the autoclave has reached a pressure of 
0.2 bar, a bolt enters the locking mechanism of the 
sliding door and prevents its opening during 
process. After the pressure drops to "0", this bolt 
releases the locking mechanism so that the sliding 
door can be unlocked by a 90° turn counter-
clockwise and be pushed back (marking on knob is 
now along with the apparatus' axis), see also 
chapter 4.5. If you try to open the door before 
pressure is 0 bar, you will not succeed because the 
bolt is still locked in. Thus, the sliding door cannot be 
completely unlocked and opened. 
In this case, lock the door once again (marking is 
now crosswise with the apparatus' axis), wait until 
pressure drops to "0" and then unlock and open the 
door. 
If the sliding door still cannot be opened, then it is 
jammed by long instruments which protrude too far 
upward -- or the mechanical system of the door is 
defective. Try to slide the door backward by applying 
slightly more force. 

If this is not possible you must notify the MELAG 
Customer Service Department. 

Pressure release takes too long 

During normal operations, the pressure indication 
has fallen to 

0“ by the time the lamp 

END“ (13) 

glows and the acoustic signal sounds. If, however, 
an extremely great number of unpacked metal 
instruments are sterilized together, then re-
evaporation of the resulting condensate can cause a 
situation in which the operator must wait a few 
seconds after the acoustic signal before the 
pressure has completely fallen and the door allows 
itself to be opened (as described above).  
When the pressure gauge shows a value of 0.2 bar 
or more at the end of a cycle (light 

END“ (13) glows, 

with a beep), there is dirt inside the apparatus which 
has settled down to the steam filter in front of the 
solenoid valve of the pressure release (41). 
This filter can be reached from the lower side of the 
autoclave without dismantling it (use a spanner with 
14 mm). 
The filter must be cleaned by a service engineer or 
replaced. During this service work, the adjacent 
openings should be rinsed and blown out before the 
autoclave is reassembled. To perform this, first 
remove the filter, then close off the outgoing-water 
hose to prevent water from flowing through it. Then 
pour demineralized or distilled water into the 
pressure chamber and allow it to run through. Then 
close the sliding door, and blow through the 
outgoing waste water hose. 

7.9 

The sliding door cannot be easily moved ____________________________________

 

Normally, the sliding door (1), when unlocked -- see 
Section 4.5 -- can be very easily moved: for 
example, with 2 fingers. This is because of the 
following: the actual metal closure floats over the 
door gasket, the door slides on PTFE bearings, and 
the door moves into a recess at the rear area of the 
autoclave. If the door, however, is difficult to move, 
please consider the following possibilities: 

Transport damage 

If the autoclave has been subjected to violent impact 
from the side during transport, it is possible that the 
suspension of the upper autoclave plastic cladding 
has been bent. In this case, it is possible that the 
collar of the cladding of the sliding door (under the 
locking handle) collides toward the rear with a 
corner of the opening in the upper part of the 
autoclave cladding. You must notify the MELAG 
Customer Service Department. The technician must 
adjust the suspension of the upper autoclave plastic 
cladding in order that the collar runs properly in the 
centre of the opening. 

The metal closure cover does not completely lift 
during unlocking. 

If the actual metal closure cover (5 mm thick) does 
not lift, or completely lift, although the door is fully 
unlocked, then the closure will not float over the 
gasket and its upper side will not lie flush at the 
upper corner of the guide slot. The operator can 
then provisionally lift the closure by pressing on the 
plastic cladding of the sliding door, and by pulling on 
the locking handle, and can then open the 
autoclave. 
In every case you must notify the MELAG Customer 
Service Department. 
The problem may be that a technician has forgotten 
to reinstall the PTFE bearing pieces between the 
closure cladding and the slide rails after a repair job. 
Or, the 4 spring-loaded hexagon socket-heat screws 
have jammed because of soiling, debris from paint, 
or a crooked position in the guides of the closure 
cladding. In such case, unscrew the screws, clean 
the guides, and screw the screws back in tight by 
hand, using the springs and a thread-locking 
adhesive. 

7.10 

Water reservoir goes empty _______________________________________________

 

If the water gets from the storage tank to the 
condenser container and thus gets empty (over 
night), there is a leak in the non-return valve of the 
pump. 
Inform the service engineer who will clean or replace 

the pump. 
The autoclave can be used in the meantime, if the 
cycle successfully runs the end: i.e., if the lamp 

END“ (13) lights up. 

Summary of Contents for MELAquick 12

Page 1: ...facturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testify to the ex ceptional quality of our products which are manufactured exclusively in Germany As all oth...

Page 2: ...parametric regulation 4 1 5 Printout of the log 5 1 6 Water one way system 5 1 7 No start of program if water quality is inadequate 5 2 Introduction 6 2 1 Preparation of instruments 6 2 2 Formation of...

Page 3: ...re 17 7 6 Excessive indication of pressure and or temperature 17 7 7 Hissing during run foggy instruments 17 7 8 Sliding door cannot be unlocked 18 7 9 The sliding door cannot be easily moved 18 7 10...

Page 4: ...rilization mode 2 Locking handle 10 Signal lamp Fault 17 Signal lamp Disinfection mode 3 Water storage tankr 11 Signal lamp Ready 18 Connector for waste water hose 4 Power switch 12 Signal lamp Cycle...

Page 5: ...Operation manual MELAquick 12 Page 2 1 1 2 Rear View______________________________________________________________________...

Page 6: ...ection steam generator 41 Steam filter before solenoid valve 49 Solenoid valve for pressure release 34 Temperature sensor pre heating 42 Safety valve 50 Micro switch for door lock 35 Temperature senso...

Page 7: ...e leaving the factory the autoclave is set to sterilization The operator should select the desired program as soon as possible after switching on the electrical power switch Power on 4 If the autoclav...

Page 8: ...not only the autoclave but also the instruments will be damaged and become dirty However in daily practice the instruments are prepared insufficiently This is critical especially for the sterilization...

Page 9: ...n are covered with residues of oil dirt or grease coming form the production 2 2 Formation of rust drag in ________________________________________________ It has already been said that the materials...

Page 10: ...to successfully verify the sterilisation effectiveness of the autoclave only in conjunction with the sterilization program Test spores which are already packed for example spore strips in paper packi...

Page 11: ...ck the water level in the container If the water reaches the MAX mark the container must be emptied immediately We recommend to empty the condenser container each time purified water is added to the r...

Page 12: ...____________________ When the MELAquick 12 is put into operation for the first time it is absolutely necessary to run two consecutive sterilization cycles with no instruments in the autoclave This is...

Page 13: ...rilze liquids or textiles Observe the instructions above on the reheating and cool down phase if the program is changed between sterilization and disinfection after the autoclave has been preheated 4...

Page 14: ...ry heating up phase When the required pressure 2 bar and temperature 134 C for sterilization 0 2 bar and 105 C for disinfection is reached the lamp RUNNING shows a continuous light which announces tha...

Page 15: ...____________________________________ Overheat excluded Overheat during error free process is generally excluded due to the concept of separate steam generation in combination with electronic parametri...

Page 16: ...__________________________ Because of its parameters control system and its optimised heating the MELAG autoclave MELAquick 12 can be used for sterilization disinfection without interruption So 8 10 s...

Page 17: ...mpair the autoclave s functional effectiveness Aggressive water pH 5 or 7 can also damage the apparatus It is only for high values of conductivity that the autoclave triggers a warning or fault alarm...

Page 18: ..._________________________________ If freezing temperatures are possible during transport any water which may remain in the feed water system must be removed in order to prevent damage to the pump and...

Page 19: ...____________________________ Preheat phase cool down phase During the following phases it is not possible to start a program During the approx 10 15 min preheating phase only once after first switchin...

Page 20: ...into the steam channel of the chamber from the bottom has to be checked for resistances on plugs and contacts may have to be cleaned or the sensor to be replaced For questions about the precision of...

Page 21: ...e chamber and allow it to run through Then close the sliding door and blow through the outgoing waste water hose 7 9 The sliding door cannot be easily moved ____________________________________ Normal...

Page 22: ...time of the fault e g during pre heat phase or during pressure generation or during sterilization and inform the service engineer in order to make a quick and low priced repair possible The autoclave...

Page 23: ...In this hose and thus in the water discharge hose you can normally see small amounts of flowing steam condensat The presence of prescribed temperature 134 C or 105 C will additionally be indicated by...

Page 24: ...water reservoir to the steam generator Fault diagnosis If this water dose is too small the pump will start again at release of the steam generator s overheat protection and after the end of the real p...

Page 25: ...on 2 3 2 If the amount of water is too small refer to Section 1 1 1 If the amount is correct optically check all joints for leakage when they are under pressure Watch the water discharge hose of the s...

Page 26: ...the clickpoint must be checked The correct clickpoint is as follows when the autoclave is preheated and when the sliding door is locked the microswitch should click approximately halfway through the...

Page 27: ...e tank with demineralized or distilled water as described in Section 5 3 Then fill again with adequate water see Section 3 3 8 7 Fault No 15____________________________________________________________...

Reviews: