Page 15
5.3.3 Refilling / Replacement __________________________________________________________
Lamp
„
WATER“, indicating lack of water
The light of lamp "WATER" shows that the amount
of water in the reservoir is not sufficient any more.
With this lamp on, there is no possibility to start a
program. The reservoir has to be refilled with aqua
dest/aqua dem until the lamp goes out. The
maximum level is reached with the lower edge of the
trapezi-form filling nozzle.
If lamp "WATER" comes on after a start of a process
and during pumping, it has no influence on running
the program.
The remaining amount of water in the water storage
tank is in any case sufficient for the started
sterilisation or disinfection cycle to run to completion.
Replacement in case of pollution
If dirt has come into the water reservoir because of
misoperation, the reservoir has to be rinsed and
cleaned, e.g. with a bottle brush. In every case, let
off the water as described above and refill the
reservoir.
6
Putting out of operation / Transport
•
Switch off main switch.
•
Pull out main plug, let autoclave cool down.
•
Remove hose to condenser container from
autoclave by pressing hard against the black
ring of the connector (16) and pulling the hose
out of the connector at the same time.
•
Empty water reservoir as described in
chapter 5.3.2.
•
If possible,
use the original packing
to
transport the autoclave. Be sure in any case to
additionally seal the autoclave inside plastic
wrapping, in order to prevent packing material
from entering the water storage tank, etc.
6.1
Transport under danger of frost ____________________________________________
If freezing temperatures are possible during
transport, any water which may remain in the feed-
water system must be removed in order to prevent
damage to the pump and the water lines. A service
technician is required for this work, which includes
the following steps:
•
The autoclave must be switched on and
preheated.
•
Empty the water storage tank as described in
Section 5.3.2.
•
Remove the metal side-enclosure panels from
the autoclave.
•
At the bottom left on the electronic circuit board
there are four wire-clasp buttons / switches (22).
The second from the bottom is the Test Button,
which sets all outputs of the control system.
Press this Test Button for the electronic control
system until no more water is pumped on the
output side of the water pump (through the thin,
transparent hose).
•
Open and close the sliding door several times
until all the water remaining in the steam
generator is expelled. This action will heat the air
contained in the preheated pressure chamber,
which will cause the air to expand and expel any
water remaining into the condensor container.
•
Switch off the power switch and unplug the
power cable.
•
Replace the metal side-enclosure panels on the
autoclave.
•
Disconnect hose to condenser container from the
autoclave by pressing hard against the black ring
of the autoclave's connector; at the same time
pull the hose.
•
Let the autoclave cool down.
7
Breakdowns without warning light
The following instructions for breakdowns are pro-
vided to help the user resolve minor trouble himself,
or to allow him a more accurate description of
malfunctions to his authorized dealer, warehouse
agent, or customer-service representative.
Bigger faults or damages are repaired only by
appointed service engineers or by the manu-
facturer.