5.3.3  Refilling / Replacement __________________________________________________________ 

Lamp 

WATER“, indicating lack of water 

The light of lamp "WATER" shows that the amount 
of water in the reservoir is not sufficient any more. 
With this lamp on, there is no possibility to start a 
program. The reservoir has to be refilled with aqua 
dest/aqua dem until the lamp goes out. The 
maximum level is reached with the lower edge of the 
trapezi-form filling nozzle. 
If lamp "WATER" comes on after a start of a process 
and during pumping, it has no influence on running 

the program. 
The remaining amount of water in the water storage 
tank is in any case sufficient for the started 
sterilisation or disinfection cycle to run to completion. 

Replacement in case of pollution 

If dirt has come into the water reservoir because of 
misoperation, the reservoir has to be rinsed and 
cleaned, e.g. with a bottle brush. In every case, let 
off the water as described above and refill the 
reservoir. 

Putting out of operation / Transport 

 

 

Switch off main switch. 

 

Pull out main plug, let autoclave cool down. 

  Remove hose to condenser container from 

autoclave by pressing hard against the black 
ring of the connector (16) and pulling the hose 
out of the connector at the same time.

  Empty water reservoir as described in 

chapter 5.3.2. 

 If possible, 

use the original packing 

to 

transport the autoclave. Be sure in any case to 
additionally seal the autoclave inside plastic 
wrapping, in order to prevent packing material 
from entering the water storage tank, etc. 

6.1 

Transport under danger of frost ____________________________________________

 

If freezing temperatures are possible during 
transport, any water which may remain in the feed-
water system must be removed in order to prevent 
damage to the pump and the water lines. A service 
technician is required for this work, which includes 
the following steps: 

  The autoclave must be switched on and 

preheated. 

  Empty the water storage tank as described in 

Section 5.3.2. 

  Remove the metal side-enclosure panels from 

the autoclave. 

 

At the bottom left on the electronic circuit board 
there are four wire-clasp buttons / switches (22). 
The second from the bottom is the Test Button, 
which sets all outputs of the control system. 
Press this Test Button for the electronic control 
system until no more water is pumped on the 
output side of the water pump (through the thin, 
transparent hose). 

  Open and close the sliding door several times 

until all the water remaining in the steam 
generator is expelled. This action will heat the air 
contained in the preheated pressure chamber, 
which will cause the air to expand and expel any 
water remaining into the condensor container. 

  Switch off the power switch and unplug the 

power cable. 

  Replace the metal side-enclosure panels on the 

autoclave. 

  Disconnect hose to condenser container from the 

autoclave by pressing hard against the black ring 
of the autoclave's connector; at the same time 
pull the hose. 

  Let the autoclave cool down. 

Breakdowns without warning light 

 

The following instructions for breakdowns are pro-
vided to help the user resolve minor trouble himself, 
or to allow him a more accurate description of 
malfunctions to his authorized dealer, warehouse 
agent, or customer-service representative.  

Bigger faults or damages are repaired only by 
appointed service engineers or by the manu-
facturer.

 

Summary of Contents for MELAquick 12

Page 1: ...facturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testify to the ex ceptional quality of our products which are manufactured exclusively in Germany As all oth...

Page 2: ...parametric regulation 4 1 5 Printout of the log 5 1 6 Water one way system 5 1 7 No start of program if water quality is inadequate 5 2 Introduction 6 2 1 Preparation of instruments 6 2 2 Formation of...

Page 3: ...re 17 7 6 Excessive indication of pressure and or temperature 17 7 7 Hissing during run foggy instruments 17 7 8 Sliding door cannot be unlocked 18 7 9 The sliding door cannot be easily moved 18 7 10...

Page 4: ...rilization mode 2 Locking handle 10 Signal lamp Fault 17 Signal lamp Disinfection mode 3 Water storage tankr 11 Signal lamp Ready 18 Connector for waste water hose 4 Power switch 12 Signal lamp Cycle...

Page 5: ...Operation manual MELAquick 12 Page 2 1 1 2 Rear View______________________________________________________________________...

Page 6: ...ection steam generator 41 Steam filter before solenoid valve 49 Solenoid valve for pressure release 34 Temperature sensor pre heating 42 Safety valve 50 Micro switch for door lock 35 Temperature senso...

Page 7: ...e leaving the factory the autoclave is set to sterilization The operator should select the desired program as soon as possible after switching on the electrical power switch Power on 4 If the autoclav...

Page 8: ...not only the autoclave but also the instruments will be damaged and become dirty However in daily practice the instruments are prepared insufficiently This is critical especially for the sterilization...

Page 9: ...n are covered with residues of oil dirt or grease coming form the production 2 2 Formation of rust drag in ________________________________________________ It has already been said that the materials...

Page 10: ...to successfully verify the sterilisation effectiveness of the autoclave only in conjunction with the sterilization program Test spores which are already packed for example spore strips in paper packi...

Page 11: ...ck the water level in the container If the water reaches the MAX mark the container must be emptied immediately We recommend to empty the condenser container each time purified water is added to the r...

Page 12: ...____________________ When the MELAquick 12 is put into operation for the first time it is absolutely necessary to run two consecutive sterilization cycles with no instruments in the autoclave This is...

Page 13: ...rilze liquids or textiles Observe the instructions above on the reheating and cool down phase if the program is changed between sterilization and disinfection after the autoclave has been preheated 4...

Page 14: ...ry heating up phase When the required pressure 2 bar and temperature 134 C for sterilization 0 2 bar and 105 C for disinfection is reached the lamp RUNNING shows a continuous light which announces tha...

Page 15: ...____________________________________ Overheat excluded Overheat during error free process is generally excluded due to the concept of separate steam generation in combination with electronic parametri...

Page 16: ...__________________________ Because of its parameters control system and its optimised heating the MELAG autoclave MELAquick 12 can be used for sterilization disinfection without interruption So 8 10 s...

Page 17: ...mpair the autoclave s functional effectiveness Aggressive water pH 5 or 7 can also damage the apparatus It is only for high values of conductivity that the autoclave triggers a warning or fault alarm...

Page 18: ..._________________________________ If freezing temperatures are possible during transport any water which may remain in the feed water system must be removed in order to prevent damage to the pump and...

Page 19: ...____________________________ Preheat phase cool down phase During the following phases it is not possible to start a program During the approx 10 15 min preheating phase only once after first switchin...

Page 20: ...into the steam channel of the chamber from the bottom has to be checked for resistances on plugs and contacts may have to be cleaned or the sensor to be replaced For questions about the precision of...

Page 21: ...e chamber and allow it to run through Then close the sliding door and blow through the outgoing waste water hose 7 9 The sliding door cannot be easily moved ____________________________________ Normal...

Page 22: ...time of the fault e g during pre heat phase or during pressure generation or during sterilization and inform the service engineer in order to make a quick and low priced repair possible The autoclave...

Page 23: ...In this hose and thus in the water discharge hose you can normally see small amounts of flowing steam condensat The presence of prescribed temperature 134 C or 105 C will additionally be indicated by...

Page 24: ...water reservoir to the steam generator Fault diagnosis If this water dose is too small the pump will start again at release of the steam generator s overheat protection and after the end of the real p...

Page 25: ...on 2 3 2 If the amount of water is too small refer to Section 1 1 1 If the amount is correct optically check all joints for leakage when they are under pressure Watch the water discharge hose of the s...

Page 26: ...the clickpoint must be checked The correct clickpoint is as follows when the autoclave is preheated and when the sliding door is locked the microswitch should click approximately halfway through the...

Page 27: ...e tank with demineralized or distilled water as described in Section 5 3 Then fill again with adequate water see Section 3 3 8 7 Fault No 15____________________________________________________________...