background image

 

 

Page 13 

4.14 

Process control _________________________________________________________

 

The autoclave offers the following possibilities for 
controlling a safe sterilization/disinfection process: 

Successful conclusion of a process 

As soon as a sterilization/disinfection is completed 
successfully, i.e. keeping the required parameters 
for pressure and temperature in the prescribed time, 
the lamp 

END“ lights and an acoustic signal can be 

heard. If the protocol printer is connected, the 
successful end of the cycle will be documented on 
the printout. 

Pressure-temperature correlation 

During process, the correlation of temperature with 
pressure can be controlled by observing the 
thermometer (6) and the pressure gauge (5), 
according to the curve for saturated steam (e.g. 2-
2.1 bar with 134-135°C). The precision of the 
indicator instruments must be taken into account 
here: see Section 5.2. 

Electronic monitoring of parameters 

During running of the program, the following 
parameters will be automatically monitored under 
microprocessor control: 
Heat-up time 
The maximum heat-up time is normally 3 minutes. 
If overloading, insufficient venting, or fault in 
operation prevent the autoclave from achieving the 
values required by the selected program for 

pressure and temperature within 5 min., then the 
program will interrupt owing to time overrun, and the 
lamps for 

FAULT“ (10) and 

RUNNING“ (12) will 

flash. 
Sterilization/disinfection time 
During sterilization/disinfection process, the 
parameters pressure and temperature are 
constantly controlled. The sterilization/disinfection 
time of 4 / 6 minutes will only run if both parameters 
reach the required values (lamp "RUNNING" lights). 
If the autoclave exceeds the monitoring time of 
5.5 minutes 

for 

sterilization / 7.5 minutes 

for 

disinfection due to faulty run, the program will also 
be stopped because of exceeding time. The lamps 
for 

FAULT“ (10) and 

RUNNING“ (12) will flash in 

this case. 
Exceeding steam temperature 
If the autoclave exceeds the steam temperature of 
140°C during sterilization or 112°C during 
disinfection due to a fault, the program will also 
interrupt in such case, and the lamp for 

FAULT“ 

(10) will glow. The lamps for 

READY“ (11) and 

RUNNING“ (12) will flash. 

Temperature limitation 

The automatic temperature limitation has already 
been described (See Section 4.13). 

4.15 

Sterilizing frequency _____________________________________________________

 

Because of its parameters control system and its 
optimised heating, the MELAG-autoclave 
MELAquick

®

12

 

 can be used for 

sterilization / disinfection without interruption. 
So 8-10 sterilizations per hour are possible. 

Further informations about sterilization ________________________________  

5.1 

Total operating time ______________________________________________________

 

Operating time 

"The operating time is the time needed for 
sterilization without taking account of loading and 
emptying." (DIN 58900) 
Due to the parameter regulation of the autoclave, 
the operating time is optimised mainly depending on 
the number of items to be sterilized and the 
temperature of the apparatus. The minimum 
operating time (small load, immediate sterilization 
before) needed for sterilization is about 5.5 minutes 

and about 7.5 minutes for disinfection. The 
maximum time (maximum load, first process after 
complete pre-heat) is about 8 

minutes for 

sterilization and 9 minutes for disinfection. 

Included times 

The operating time includes times for pressure 
generation, ventilation and heat-up of instruments 
and chamber as well as for sterilization which 
consists of equalization time, killing time and an 
extra charge for safety. 

5.2 

Accuracy of process regulation ____________________________________________

 

The analogue instruments for pressure and 
temperature of the MELAquick

®

12

 

 work completely 

independently from the microprocessor regulation 
and are only for additional information about the run 
state of the autoclave. 
Because of their small size, the accuracy of these 
instruments is only 

±

 0,1 bar (pressure gauge) and  

±

 2°C (thermometer). 

Owing to the time delay of the temperature sensor, it 
can happen that the indication of the temperature 
reaches the correct value only after beginning of the 
sterilization or disinfection time: i.e., while the lamp 

RUNNING“ (12) is glowing, and after the required 

temperature has already been reaches in the 

chamber. 
It may also happen during continuous operation, as 
a result of constant high temperature of the 
autoclave, that the temperature indicates on the 
thermometer is 2°C higher than the temperature 
shown for the first sterilization or disinfection cycle. 
In the actual sterilization compartment, however, 
there will be no temperature deviation. 

Microprocessor control 

The electronic control system, on the other hand, 
operates with an absolute precision of 0.5°C and 
controls the sterilization or disinfection temperature 
with a maximum deviation of 

±

 0.3°C. 

The microprocessor regulation in combination with 

Summary of Contents for MELAquick 12

Page 1: ...facturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testify to the ex ceptional quality of our products which are manufactured exclusively in Germany As all oth...

Page 2: ...parametric regulation 4 1 5 Printout of the log 5 1 6 Water one way system 5 1 7 No start of program if water quality is inadequate 5 2 Introduction 6 2 1 Preparation of instruments 6 2 2 Formation of...

Page 3: ...re 17 7 6 Excessive indication of pressure and or temperature 17 7 7 Hissing during run foggy instruments 17 7 8 Sliding door cannot be unlocked 18 7 9 The sliding door cannot be easily moved 18 7 10...

Page 4: ...rilization mode 2 Locking handle 10 Signal lamp Fault 17 Signal lamp Disinfection mode 3 Water storage tankr 11 Signal lamp Ready 18 Connector for waste water hose 4 Power switch 12 Signal lamp Cycle...

Page 5: ...Operation manual MELAquick 12 Page 2 1 1 2 Rear View______________________________________________________________________...

Page 6: ...ection steam generator 41 Steam filter before solenoid valve 49 Solenoid valve for pressure release 34 Temperature sensor pre heating 42 Safety valve 50 Micro switch for door lock 35 Temperature senso...

Page 7: ...e leaving the factory the autoclave is set to sterilization The operator should select the desired program as soon as possible after switching on the electrical power switch Power on 4 If the autoclav...

Page 8: ...not only the autoclave but also the instruments will be damaged and become dirty However in daily practice the instruments are prepared insufficiently This is critical especially for the sterilization...

Page 9: ...n are covered with residues of oil dirt or grease coming form the production 2 2 Formation of rust drag in ________________________________________________ It has already been said that the materials...

Page 10: ...to successfully verify the sterilisation effectiveness of the autoclave only in conjunction with the sterilization program Test spores which are already packed for example spore strips in paper packi...

Page 11: ...ck the water level in the container If the water reaches the MAX mark the container must be emptied immediately We recommend to empty the condenser container each time purified water is added to the r...

Page 12: ...____________________ When the MELAquick 12 is put into operation for the first time it is absolutely necessary to run two consecutive sterilization cycles with no instruments in the autoclave This is...

Page 13: ...rilze liquids or textiles Observe the instructions above on the reheating and cool down phase if the program is changed between sterilization and disinfection after the autoclave has been preheated 4...

Page 14: ...ry heating up phase When the required pressure 2 bar and temperature 134 C for sterilization 0 2 bar and 105 C for disinfection is reached the lamp RUNNING shows a continuous light which announces tha...

Page 15: ...____________________________________ Overheat excluded Overheat during error free process is generally excluded due to the concept of separate steam generation in combination with electronic parametri...

Page 16: ...__________________________ Because of its parameters control system and its optimised heating the MELAG autoclave MELAquick 12 can be used for sterilization disinfection without interruption So 8 10 s...

Page 17: ...mpair the autoclave s functional effectiveness Aggressive water pH 5 or 7 can also damage the apparatus It is only for high values of conductivity that the autoclave triggers a warning or fault alarm...

Page 18: ..._________________________________ If freezing temperatures are possible during transport any water which may remain in the feed water system must be removed in order to prevent damage to the pump and...

Page 19: ...____________________________ Preheat phase cool down phase During the following phases it is not possible to start a program During the approx 10 15 min preheating phase only once after first switchin...

Page 20: ...into the steam channel of the chamber from the bottom has to be checked for resistances on plugs and contacts may have to be cleaned or the sensor to be replaced For questions about the precision of...

Page 21: ...e chamber and allow it to run through Then close the sliding door and blow through the outgoing waste water hose 7 9 The sliding door cannot be easily moved ____________________________________ Normal...

Page 22: ...time of the fault e g during pre heat phase or during pressure generation or during sterilization and inform the service engineer in order to make a quick and low priced repair possible The autoclave...

Page 23: ...In this hose and thus in the water discharge hose you can normally see small amounts of flowing steam condensat The presence of prescribed temperature 134 C or 105 C will additionally be indicated by...

Page 24: ...water reservoir to the steam generator Fault diagnosis If this water dose is too small the pump will start again at release of the steam generator s overheat protection and after the end of the real p...

Page 25: ...on 2 3 2 If the amount of water is too small refer to Section 1 1 1 If the amount is correct optically check all joints for leakage when they are under pressure Watch the water discharge hose of the s...

Page 26: ...the clickpoint must be checked The correct clickpoint is as follows when the autoclave is preheated and when the sliding door is locked the microswitch should click approximately halfway through the...

Page 27: ...e tank with demineralized or distilled water as described in Section 5 3 Then fill again with adequate water see Section 3 3 8 7 Fault No 15____________________________________________________________...

Reviews: