background image

 

Operation Manual 

MELA

quick

®

12

 

Page 20 

8.2 

Fault No. 4 ______________________________________________________________

 

Cause of fault: 

The autoclave did not achieve its parameter for 
pressure and/or temperature within the supervision 
time of 5.5 minutes. 

Time of fault: 

After program start, during pressure generation 
(lamp "READY" blinks). 

The following reasons can be given:

 

8.2.1  Autoclave insufficiently ventilated _________________________________________________ 

In this case, the feed water pump runs only once 
after the start of the program. 

For proper function of the autoclave (death of 
germs), it is necessary to extract air, respectively 
air/steam-mixture, out of the chamber. After a 
program start or during pressure generation, air is 
extracted through a current channel which branches 
off in the rear part of the chamber, leads to the 
measuring block and then thron a fine filter to the 
inside flow nozzle (43). In this channel the steam-
temperature sensor for the electronic system (35) 
(visable from above in the rear opening of the 
chamber bottom, under the sieve insert for the 
chamber), as well as the temperature sensor of the 
thermometer (39), are contacted by the flowing 
steam. In front of the hose connection to the 
measuring block is an additional steam filter 
screwed inside the chamber block (44). 

Fault diagnosis: 

The autoclave will not properly vent if any of the 
following components are clogged: the sieve insert 
for the pressure chamber, the fine filter installed 
upstream of the nozzle (43), or the steam filter (44). 
The autoclave does achieve a pressure of approx. 
2.2 

bar (for sterilization), or 0.3--0.4 

bar for 

disinfection within a period of 1.5...3 min. after the 
start of the program. Although it maintains this 
pressure, an excessively

low temperature (or no temperature at all) is 
indicated on the thermometer. The autoclave then 
switches over to fault no. 4 after program start, since 
the partial pressure of the air is added to that of the 
steam, which prevents the saturated-steam 
temperature from being achieved at the prescribed 
pressure. 
Call for a service engineer. 
Beyond, every blockage in this area can be 
recognised with the transparent Teflon hose which 
leads from the measuring block to a T-junction. In 
this hose (and thus in the water discharge hose), 
you can normally see small amounts of flowing 
steam/condensat. The presence of prescribed 
temperature (134°C or 105°C) will additionally be 
indicated by a LED on the electronic CPU-board. 

Fault correction: 

Remove the sieve insert. Optical control of 
ventilation hole in the chamber (bottom of chamber 
at rear rim). This hole should not be covered or 
blocked. Check whether somebody has worked 
without sieve insert so that instruments have 
covered the ventilation hole. 
Check flow nozzle (43) and filters (40) and (44), if 
necessary also steam filter (34), probably clean or 
replace them. Before complete re-assembly it might 
be useful to run the autoclave once without fine filter 
(40) and flow nozzle (43) to flush the tubing system 
and remove any residues. 

8.2.2  Pressure switch is wrongly adjusted/defective _______________________________________ 

In this case, the feed water pump runs only once 
after the start of the program. 
Diagnosis of fault: 

If the autoclave achieves a pressure for sterilization 
of less than 2.1 bar (or 0.2 bar for disinfection) within 
1.5 ... 3 

min., if it remains at this insufficient 

pressure, and if the thermometer reaches the 
corresponding saturated-steam temperature, then 
the pressure switch for sterilization (53) or for 
disinfection (54) is not properly set -- or, it is 
defective. 
The autoclave indicates fault no. 4 without a 
repeated start of the pump because the prescribed 
supervision temperature cannot be achieved. 
Call for a service engineer. 

Fault correction: 

Adjust the respective pressure switch by turning the 
justification screw clockwise so that during pressure 
generation, a pressure of 2.25...2.3 

bar 

(0.3...0.4 bar) will be reached. If the adjustment has 
no success, replace the pressure switch. 
The autoclave must be placed in the pure pressure-
control mode in order to adjust the pressure switch. 
To make this adjustment, first switch off the power 
switch (4). Then hook in the two upper wire-clasp 
switches (A and B) of the 4 wire-clasp buttons / 
switches (32) on the electronic circuit board, at the 
lower left on the circuit board (on a blue plastic 
base). The second switch from the top has already 
been hooked in at the factory. After adjustment work 
is completed, make sure that the power switch is off, 
and unhook the upper wire-clasp switch (A). If 
adjustment is not possible, the pressure switch must 
be replaced, and the new switch must be properly 
adjusted. 

Summary of Contents for MELAquick 12

Page 1: ...facturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testify to the ex ceptional quality of our products which are manufactured exclusively in Germany As all oth...

Page 2: ...parametric regulation 4 1 5 Printout of the log 5 1 6 Water one way system 5 1 7 No start of program if water quality is inadequate 5 2 Introduction 6 2 1 Preparation of instruments 6 2 2 Formation of...

Page 3: ...re 17 7 6 Excessive indication of pressure and or temperature 17 7 7 Hissing during run foggy instruments 17 7 8 Sliding door cannot be unlocked 18 7 9 The sliding door cannot be easily moved 18 7 10...

Page 4: ...rilization mode 2 Locking handle 10 Signal lamp Fault 17 Signal lamp Disinfection mode 3 Water storage tankr 11 Signal lamp Ready 18 Connector for waste water hose 4 Power switch 12 Signal lamp Cycle...

Page 5: ...Operation manual MELAquick 12 Page 2 1 1 2 Rear View______________________________________________________________________...

Page 6: ...ection steam generator 41 Steam filter before solenoid valve 49 Solenoid valve for pressure release 34 Temperature sensor pre heating 42 Safety valve 50 Micro switch for door lock 35 Temperature senso...

Page 7: ...e leaving the factory the autoclave is set to sterilization The operator should select the desired program as soon as possible after switching on the electrical power switch Power on 4 If the autoclav...

Page 8: ...not only the autoclave but also the instruments will be damaged and become dirty However in daily practice the instruments are prepared insufficiently This is critical especially for the sterilization...

Page 9: ...n are covered with residues of oil dirt or grease coming form the production 2 2 Formation of rust drag in ________________________________________________ It has already been said that the materials...

Page 10: ...to successfully verify the sterilisation effectiveness of the autoclave only in conjunction with the sterilization program Test spores which are already packed for example spore strips in paper packi...

Page 11: ...ck the water level in the container If the water reaches the MAX mark the container must be emptied immediately We recommend to empty the condenser container each time purified water is added to the r...

Page 12: ...____________________ When the MELAquick 12 is put into operation for the first time it is absolutely necessary to run two consecutive sterilization cycles with no instruments in the autoclave This is...

Page 13: ...rilze liquids or textiles Observe the instructions above on the reheating and cool down phase if the program is changed between sterilization and disinfection after the autoclave has been preheated 4...

Page 14: ...ry heating up phase When the required pressure 2 bar and temperature 134 C for sterilization 0 2 bar and 105 C for disinfection is reached the lamp RUNNING shows a continuous light which announces tha...

Page 15: ...____________________________________ Overheat excluded Overheat during error free process is generally excluded due to the concept of separate steam generation in combination with electronic parametri...

Page 16: ...__________________________ Because of its parameters control system and its optimised heating the MELAG autoclave MELAquick 12 can be used for sterilization disinfection without interruption So 8 10 s...

Page 17: ...mpair the autoclave s functional effectiveness Aggressive water pH 5 or 7 can also damage the apparatus It is only for high values of conductivity that the autoclave triggers a warning or fault alarm...

Page 18: ..._________________________________ If freezing temperatures are possible during transport any water which may remain in the feed water system must be removed in order to prevent damage to the pump and...

Page 19: ...____________________________ Preheat phase cool down phase During the following phases it is not possible to start a program During the approx 10 15 min preheating phase only once after first switchin...

Page 20: ...into the steam channel of the chamber from the bottom has to be checked for resistances on plugs and contacts may have to be cleaned or the sensor to be replaced For questions about the precision of...

Page 21: ...e chamber and allow it to run through Then close the sliding door and blow through the outgoing waste water hose 7 9 The sliding door cannot be easily moved ____________________________________ Normal...

Page 22: ...time of the fault e g during pre heat phase or during pressure generation or during sterilization and inform the service engineer in order to make a quick and low priced repair possible The autoclave...

Page 23: ...In this hose and thus in the water discharge hose you can normally see small amounts of flowing steam condensat The presence of prescribed temperature 134 C or 105 C will additionally be indicated by...

Page 24: ...water reservoir to the steam generator Fault diagnosis If this water dose is too small the pump will start again at release of the steam generator s overheat protection and after the end of the real p...

Page 25: ...on 2 3 2 If the amount of water is too small refer to Section 1 1 1 If the amount is correct optically check all joints for leakage when they are under pressure Watch the water discharge hose of the s...

Page 26: ...the clickpoint must be checked The correct clickpoint is as follows when the autoclave is preheated and when the sliding door is locked the microswitch should click approximately halfway through the...

Page 27: ...e tank with demineralized or distilled water as described in Section 5 3 Then fill again with adequate water see Section 3 3 8 7 Fault No 15____________________________________________________________...

Reviews: