34 / 54
9018116
The control light above the cleaning key lights up. The cleaning and disinfection
machine cleans and disinfects automatically and switches off the wash pro-
gramme after completion.
The cleaning and disinfection time may differ from the set programme cycle time if
the programme cycle time is not sufficient to heat up the boiler and tank water to
the preset temperature. In this case, the cleaning cycle time is automatically ex-
tended. This means that the cleaning and disinfection machine runs until the re-
quired temperatures are reached, but max. 5 minutes.
The cleaning and disinfection programme starts and automatically runs
through to the program end.
During the programme run, the wash chamber door is locked.
9.3.12 Removal of regulator or combi-rack from the compressed air pipe
Press the locking button to disconnect the com-
pressed air pipe.
Press the locking button again and pull the com-
pressed air hose out of the coupling.
9.3.13 Removing the washware
Caution
Danger of burns and scalding due to hot clean water, washware and machine
parts
Contact with hot clean water, washware and machine parts can result in
burns/scalding of the skin.
• Wear protective gloves if necessary.
• Let the washware cool down before emptying, if necessary.
• Let the machine cool down before touching machine parts, if necessary.
• Never open the machine door or hood during a wash cycle.
• Only open and close the hood/door using the designated handle(s).
After the programme ends, the control light above the cleaning key turns off and
an acoustic signal sounds.
1. Open door.
2. Carefully remove the rack.
Check the disassembled parts of respirator masks and regulators for: cleanli-
ness, changes, wear (according to the manufacturer's instructions for use).
Summary of Contents for TopClean M