20 / 54
9018116
Note
Where discharge currents exceed 10 mA, the use of an RCD/FI is not recom-
mended. False alarms restrict machine availability.
MEIKO expressly excludes any liability for damages caused by improper connection
of the machine.
This includes any work in connection with services due to said connection, e.g. pro-
cessing complaints or claims due to:
• Triggered RCD/FI
• Automatic shut-down of the supply when consistency is lost in the protective
earth conductor (EN 60204-1 Chap. 8.2.8.c)
Local mains isolator
The power supply must be fused in accordance with the regulations and a lockable
mains isolator provided in the local electrical installation.
If an unearthed neutral conductor (N) is used with three-phase current, the mains
isolator must have 4 poles (with alternating current 2 poles).
If the neutral wire (N) is not earthed, you must use a 4-pole mains switch. Mains
power cables must be oil-resistant, shielded, flexible cables no lighter than an
H07RN-F cable.
Local fuse protection
The machine must be fused appropriately for the rated current shown on the de-
scriptive plaque. The electrical connection must be fused as a separately fused cir-
cuit (final circuit). Take note of the connection variant.
The requirements for limiting voltage changes, fluctuations and flicker in accordance
with DIN EN 61000-3-11 for the TopClean 60 M2 cleaning and disinfection machine
are fulfilled if the network has a current-
carrying capacity of ≥ 100 A.
Protective equipotential bonding
The screw for the potential equalisation
is located on the back of the machine
near the media connections.
The potential equalisation connection must be carried out in accordance with the
requirements of the local electricity supply companies and all applicable local regula-
tions (in Germany VDE 0100 Part 540). Incorporate the machine and any conductive
substructures and table systems into the local potential equalisation system.