Maintenance and care
60
Proceed as follows to check fuse F2 and to replace it if necessary:
Step Action
1
Loosen the screw from the cover.
2
Remove the fuse and check it for integrity.
3
Insert the fuse or, if necessary, an included replacement fuse with the same
characteristics (1.25 A/500 V H type, fast-blow; 6.3 x 32 mm)
4
Secure the cover again with the corresponding screw.
If the fuses trip again, please contact an authorised service shop to fix the problem.