Configuration
45
6
Configuration
6.1
Defining basic settings
Using the
Settings
Basic settings
menu item, the following basic settings and
adjustments can be defined:
Parameter
Description
Status
Device name
Device name, which can also be modified when necessary
Serial number
Serial number of the device, which should also be provided
when contacting customer service
License key
The license key entered for activation of the "Fault location"
option.
Tools
Reboot SFC250-SM
Forces a restart of the supply module
Load factory
settings
Resets all settings back to the original factory settings
Calibrate touch
screen
Starts the procedure for touch-screen calibration (see
page 61)
Firmware update
SFC250-PM
Starts a manual software update of the connected power
module. Such an update may also help when the status
indicator
on the power module does not go on after
starting.
Causes for these kinds of problems could be a failed software
update of the power module or mixing of modules with
different software versions.
The files needed for the update are loaded from the internal
memory of the supply module and do not need to be provided
by a USB flash drive. After pressing the button, the update
starts immediately and is indicated using the following symbol
in the main screen:
The installation may take several minutes. Do not
disconnect the supply module from the supply voltage
or the power module during installation!
Import / Export VPN
configuration
Import / Export of a VPN configuration (see page 51)