Unit Maintenance
MPQ2000-UG-EN Rev 8 October 2018
85
5.
Carefully lift on the front panel handle to lift the front panel from its enclosure,
as shown below. (Do not pass 90 degrees. If so, the current PCB ribbon cable
will disconnect.)
6.
Remove the 3 screws holding the battery cover in place.
7.
Remove the battery cover.
8.
Grasp the wires between the battery and the connector and pull straight up until
the connector unplugs from the MPQ2000.
9.
Connect a new battery to the MPQ2000 by aligning the battery connector to the
pins and pressing it in place.
10.
Place the battery in the battery compartment with the wires under it.