45
Figure 25: Report Configuration
Here the user can configure the report to contain the ratio and phase error tables and graphs for any
combination of taps or all the taps. If the CT is an ANSI metering class CT, then the accuracy
parallelograms will also be available. Place a check mark in the box beside the desired data and the
desired tap and data for that tap will be included in the report. Please note that the nameplate data can
be entered at any time either before the test or after the test has been completed and saved and the “
Ratio & Phase Error “ button will become available within the report. Also please note that by default,
only the external tap is initially included.
4.14 Excitation Curve Data
Data points for the excitation curve graphs are saved within each test file. Up to 30 data points for each
excitation curve can be displayed within the report by selecting the “Display Saturation Test Data” check
box within the report.
Additional data points can be viewed from the Saturation/Excitation Test Screen.