THERAPY SCREENS (Continued)
18. Manual Control
Use the Manual Control feature to circulate water at a set
temperature through the ArcticGel™ pads for a set
duration of time. Manual Control is not recommended for
patient temperature management. The operator is advised
to use the automatic therapy modes (e.g. Control Patient,
Cooling, Rewarming) for automatic patient temperature
monitoring and control.
NOTE:
The system patient temperature limit alarms will not
automatically stop Manual Control but will only alarm the user
to an over limit condition.
Manual Control
can be disabled or set to the default Manual
Control water temperature and maximum time duration in the
Normothermia
Settings
or
Hypothermia
Settings
screen. If
the Manual Control feature is disabled, the
Manual Control
button is not visible on the therapy screen. When activating
Manual Control from the therapy screen, the user will have the
option to select the Manual Control time duration. The Manual
Control time duration range that is selectable will be limited by
the maximum time duration set in Manual Control settings. See
VI. Operation Guide–Therapy Settings–Manual Control
.
INITIATE THERAPY
Initiate Normothermia (Control Patient)
Normothermia therapy is initiated and managed, and patient
temperature is automatically controlled to a set target
temperature from the
Control Patient
window in the
Normothermia
therapy screen. The
Control Patient
window
displays the patient target temperature and the duration since
the initiation of normothermia therapy.
To initiate Normothermia therapy:
1) From the patient
Therapy Selection
screen, press the
button next to Normothermia to display the
Normothermia
therapy screen.
2) The default patient target temperature will display in the
Control
window towards the bottom of the screen.
3) To modify the patient target temperature, press the
Adjust
button to display the
Control Patient–Adjust
window.
4)
Control Patient
to: Use the Up and Down arrows to set
the desired target temperature to control the patient.
5)
Rewarm at a Rate
of: Use the Up and Down arrows on
the right of the screen to set the rewarming rate.
6) Press the
Save
button to save the new settings and close
the
Control Patient–Adjust
window.
7) Press
Start
, in the
Control
window to initiate therapy. You
will hear a tone and then a voice stating “Therapy Started”.
NOTE:
For patient target temperature between 32°C and
32.9°C (89.6°F to 91.2°F) to be available for selection,
Control Strategy 3 must be chosen in
Normothermia
Settings
. See
VI. Operation Guide–Therapy Settings–
Control Strategy
for further instructions on setting the
control strategy.
Initiate Hypothermia (Cooling and Rewarming)
Hypothermia therapy is initiated and managed, and patient
temperature is automatically controlled to a set target
temperature from the
Cooling
and
Rewarming
windows in the
Hypothermia
therapy screen.
The
Cooling
window displays the cooling phase patient target
temperature and the length of time remaining in the cooling
phase of the Hypothermia therapy.
The
Rewarming
window displays the rewarming phase patient
target temperature and the length of time remaining in the
rewarming phase of the Hypothermia therapy.
To initiate hypothermia therapy:
From the patient
Therapy Selection
screen, press the button
next to Hypothermia to display the
Hypothermia
therapy
screen.
1)
Cooling Settings
•
Press the
Adjust
button on the
Cooling
window to display
the
Cooling–Adjust
window. Use the Up and Down
arrows to set the desired patient target temperature and
the desired cooling duration to cool the patient before
rewarming begins.
•
Press the
Save
button to save the new settings and close
the
Cooling–Adjust
window.
•
Press the
Cancel
button to close the
Cooling–Adjust
window without saving the new settings.
NOTE:
For patient target temperature between 32.0°C
and 32.9°C (89.6°F to 91.2°F) to be available for selection,
control strategy 3 must be chosen in
Hypothermia
Settings
. See
VI. Operation Guide–Therapy Settings–
Control Strategy
for further instructions on setting the
control strategy.
2)
Rewarming Settings
•
Press the
Adjust
button on the
Rewarming
window to
display the
Rewarming–Adjust
screen. Use the Up and
Down arrows to set the desired final patient target
temperature and the desired rewarming rate.
•
Press the
Save
button to save the new settings and close
the
Rewarming–Adjust
window.
•
Press the
Cancel
button to close the
Rewarming–Adjust
window without saving the new settings.
3)
Initiate Patient Cooling
Press
Start
, in the
Cooling
window to initiate therapy. You
will hear a tone and then a voice stating “Therapy Started.”