13
site for increased patient safety when patient temperature is
not continuously monitored by a second device.
NOTE
: Patient temperature is not controlled from the Temp In
2 connector. It is for patient temperature monitoring only.
SYSTEM COMPONENTS (Continued)
Patient Temperature Output Cables
The Arctic Sun™ Temperature Management System Control
Module has the capability to output the current Temp In 1
reading to a YSI 400 compatible hospital monitor. The Arctic
Sun™ Temperature Management System temperature output
cables are available with several connector styles (e.g. Phillips,
GE, Rusch, Bard, Nellcor) for compatibility with the various
hospital monitor input cables. Select the patient temperature
cable with the correct connector style for your hospital monitor.
NOTE:
The temperatures displayed on the Arctic Sun™
Temperature Management System Control Panel and the
hospital monitor represents the same probe reading but may
not be identical due to calibration differences between the
Control Module and the monitor.
Figure IV-3.
Temperature output cables