Operator’s Manual 

14 

 

 

V.  SETUP PROCEDURE

 

 

UNPACK 

 
1)  Unpack the Arctic Sun™ Temperature Management 

System

 

Control Module and accessories. 

2)

 

Allow the Control Module to remain upright for at least two 
hours prior to completing the installation and setup 
procedure in order to allow the chiller oil to settle. Damage 
to the chiller compressor may result otherwise.

 

 

CONNECTIONS 

 
1)  Use only approved cables and accessories with the Arctic 

Sun™ Temperature Management System Control Module 
(Appendix B).  

2)  Inspect the accessories for wear, breakage, or fraying 

before use. Replace if necessary.  

3)  Connect the Fluid Delivery Line, Temp In 1 cable, Temp In 

2 cable (optional), Temp Out cable and Fill Tube to the 
back of the Control Module. 

4)  Plug the Power Cord into the wall outlet. Position the 

Arctic Sun™ Temperature Management System so that 
access to the power cord is not restricted. 

 

POWER ON 

 
1)  Power On by activating the power switch on the back of 

the device. 

2)  The patient 

Therapy Selection 

screen will appear on the 

control panel. 

 

FILL RESERVOIR 

 
Approximately four liters of sterile water will be required to fill 
the reservoir at initial installation. Fill the reservoir with sterile 
water only. When filling the Control Module during initial 
installation or when completely empty, add one vial of Arctic 
Sun™ Temperature Management System Arctic Sun Cleaning 
Solution to the sterile water. It is recommended to add the vial 
when filling with the second liter of sterile water.  

 

1)  From the patient 

Therapy Selection 

screen, select the 

button next to either Normothermia or Hypothermia under 
the New Patient heading. Select any pad type to continue. 

2)  From the 

Hypothermia 

or 

Normothermia 

therapy screen, 

press the 

Fill Reservoir 

button.  

3)  The 

Fill Reservoir 

screen will appear. Follow the 

directions on the screen. 

4)  The filling process will automatically stop when the 

reservoir is full. Continue to replace the bottles of sterile 
water until the filling process stops.  

5)  When the 

Fill Reservoir

 process is complete, the screen 

will close. 

6)  To stop the process early, press the 

Stop

 button. 

 

NOTE:

 If the filling cycle is stopped prior to completion, 

the reservoir will not be full and may requiring filling after 
fewer patient therapies have been performed.  

 

7)  Press the 

Cancel

 button to close the screen. 

 

 
 

FUNCTIONAL VERIFICATION 

 
Certificates of Conformance for calibration, performance, and 
electrical safety tests are included with the shipment of each 
Arctic Sun™ Temperature Management System. To verify the 
system will heat and cool properly, perform the following 
functional verification procedure after initial setup and 
installation of the Control Module. 
 
1)  Power On the Control Module.  
2)  From the patient 

Therapy Selection 

screen, press the 

button next to Hypothermia to display the 

Hypothermia 

therapy screen. Select the adult pad type to continue.  

3)  From the 

Hypothermia 

therapy screen, press the 

Manual 

Control 

button to open the 

Manual Control 

window. If the 

Manual Control button is not visible, select the 

Adjust

 

button at the bottom of the screen and select the 

More

 

button to access the therapy settings screen. Enable 
Manual Control and save your settings to close the 
window.  

4)  Use the Up and Down arrows to set the Manual Control

 

water target temperature to 40°C and the duration to 30 
minutes. 

5)  Press the 

Start 

button to initiate Manual Control. Allow at 

least 3 minutes for the system to stabilize.  

6)  Monitor the flow rate and water temperature in the 

System 

status area on the 

Hypothermia 

therapy screen.  

7)  Verify that the flow rate reaches at least 1.5 liters/minute.  
8)  Verify that the water temperature increases to at least 

30°C.  

9)  Press the 

Stop 

button.  

10)  Set the Manual Control water target temperature to 4°C 

and the duration to 30 minutes.  

11)  Press the 

Start 

button to initiate 

Manual Control

.  

12)  Monitor the flow rate and water temperature in the 

System 

status area of the 

Hypothermia 

therapy screen. Verify 

that the water temperature drops to at least 6°C.  

13)  Press the 

Stop 

button to stop 

Manual Control 

 

14)  Press the 

Cancel 

button to close the 

Manual Control 

window. 

15)  Before placing the device into clinical use, it is 

recommended to disable Manual Control unless requested 
by clinical users. Select the 

Adjust

 button at the bottom of 

the therapy screen and select the 

More

 button to access 

the therapy settings screen. Disable Manual Control and 
save your settings to close the window.  

16)  Power Off the Control Module. 
 

 

 

 

 

Summary of Contents for BD Arctic Sun 6000

Page 1: ...14868...

Page 2: ...Operator s Manual 2 tube...

Page 3: ...END THERAPY 24 THERAPY SETTINGS 25 ADVANCED SETUP 29 VII ALARMS 32 MEDIUM PRIORITY ALARMS 32 LOW PRIORITY ALARMS 32 VIII TROUBLESHOOTING 33 WATER 33 Water Not Cooling 33 Water Not Warming 33 Extended...

Page 4: ...o be supplied Catalogue number Indicates the manufacturer s catalogue number so that the medical device can be identified ISO 15223 1 Reference no 5 1 7 Medical Devices Symbols to be used with medical...

Page 5: ...t the marked item can be hot and should not be touched without taking care IEC TR 60878 Reference no 0029 Graphical Symbols for Electrical Equipment in Medical Practice Draining emptying To indicate t...

Page 6: ...p On transport packaging To indicate the correct upright position IEC TR 60878 Reference no 0621 Graphical Symbols for Electrical Equipment in Medical Practice Fragile handle with care On transport pa...

Page 7: ...t plug To identify a port or plug as meeting the generic requirements of the Universal Serial Bus USB To indicate that the device is plugged into a USB port or is compatible with a USB port IEC 60417...

Page 8: ...during use of the device WARNING Informs the reader about a situation which if not avoided could result in death or serious injury It may also describe potential serious adverse reactions and safety...

Page 9: ...recommended to use the following settings Water Temperature High Limit 40 C 104 F Water Temperature Low Limit 10 C 50 F Control Strategy 2 It is recommended to use the Patient Temperature High and Pat...

Page 10: ...sider increasing minimum water temperature modifying target temperature to an attainable setting or discontinuing treatment For patient warming consider decreasing maximum water temperature modifying...

Page 11: ...ctions Disposal Upon end of life dispose of in accordance with local WEEE regulations or contact your local BD Supplier or Distributor to arrange for disposal Parameter Specification Therapy Modes Nor...

Page 12: ...terile Water This is done to suppress microorganism growth in the water reservoir and hydraulic circuit For information regarding safe handling please refer to http www medivance com manuals for the C...

Page 13: ...rrent Temp In 1 reading to a YSI 400 compatible hospital monitor The Arctic Sun Temperature Management System temperature output cables are available with several connector styles e g Phillips GE Rusc...

Page 14: ...ior to completion the reservoir will not be full and may requiring filling after fewer patient therapies have been performed 7 Press the Cancel button to close the screen FUNCTIONAL VERIFICATION Certi...

Page 15: ...the Therapy Selection screen to initiate a New Patient Continue a Current Patient or access the Advanced Setup screen New Patient Normothermia If the therapy goal is to maintain a patient temperature...

Page 16: ...t Temperature B Patient Temperature 2 if enabled 6 System Status C Water Temperature D Flow Rate 7 Therapy Graph E Patient Temperature F Water Temperature G Date and Time H Progress Bar 8 Empty Pads 9...

Page 17: ...erature Units if enabled 5 Patient Status C Patient Temperature D Patient Temperature 2 if enabled 6 System Status E Water Temperature F Flow Rate 7 Therapy Graph I Patient Temperature J Water Tempera...

Page 18: ...ower down is not captured in the Event Log 3 Notification Notifications are active when attention to the system is required e g preventative maintenance When a notification is active a small icon will...

Page 19: ...mperature graph is divided into 5 segments denoted by time markers 00 00 04 00 08 00 12 00 16 00 and 20 00 During Hypothermia therapy the graph background is color coded to represent the current time...

Page 20: ...Control Strategy 1 2 3 Range 33 5 C to 38 5 C 92 3 F to 101 3 F 32 5 C to 38 5 C 90 5 F to 101 3 F Increments 0 1 C F 0 1 C F Lower limit Control Strategy 1 2 3 Range 33 0 C to 38 0 C 91 4 F to 100 4...

Page 21: ...been changed the Arctic Sun Work to Cool Trend graph will be greyed and the icon will display a grey hour glass for a period of 15 minutes 12 Control Settings Normothermia Screen The default patient...

Page 22: ...end of the range is limited by the current target temperature Rewarm at a Rate of Use the Up and Down arrows in the center of the screen to set the rewarming rate Range 0 01 to 0 50 C hour 0 02 to 0 9...

Page 23: ...he Save button to save the new settings and close the Control Patient Adjust window 7 Press Start in the Control window to initiate therapy You will hear a tone and then a voice stating Therapy Starte...

Page 24: ...itiated when the patient reaches the transport destination 9 Plug the Control Module power cord into the wall 10 Power On 11 Connect the patient temperature probes to the Patient Temperature Cables 12...

Page 25: ...temperature From the Hypothermia Settings screen press the button to the right of the Cooling Begins parameter Select the point that the therapy timer begins Range Immediately At Target Press the Save...

Page 26: ...ows to select the Manual Control maximum time duration The selected time will be the default time displayed in the Manual Control pop up window Range 15 minutes to 99 hours Increments 15 minutes from...

Page 27: ...Control Strategy parameter Use the Up and Down arrows to select the control strategy setting Range 1 2 3 Press the Save button to save the new setting and close the window Press the Cancel button to...

Page 28: ...and close the window Press the Cancel button to close the window without saving the new settings 14 Patient Temperature 2 Use the Patient Temperature 2 setting to enable or disable the Patient Temper...

Page 29: ...when the process is complete by changing the color of the patient data file text to gray Remove the flash drive from the USB port The data is saved as a text file to the flash drive s main directory...

Page 30: ...as indicated on the display screen See the Arctic Sun Temperature Management System Service Manual for calibration requirements and instructions Calibration should be done only by trained service per...

Page 31: ...ate format to the local requirements From the Advanced Setup screen press the button to the right of the Date Format parameter Use the Up and Down arrows to select the date format Range MM DD YY DD MM...

Page 32: ...l sound All Alarm settings are maintained in the event of a power interruption Non Recoverable Alarms If an Alarm condition occurs that prevents proper use of the device or proper patient treatment su...

Page 33: ...he Arctic Sun Temperature Management System therapy graph to properly assess the progress of the temperature management therapy Additionally the Arctic Sun Temperature Management System incorporates a...

Page 34: ...leshooting Water Low Water Flow for troubleshooting instructions Water temperature is responding appropriately to control patient temperature towards the target If not see VIII Troubleshooting Water W...

Page 35: ...rapid change in core temperature that the system is unable to counteract immediately No Patient Temperature Display Confirm that the patient temperature probe is properly placed in the patient Confir...

Page 36: ...m immediately following cleaning Follow all manufacturers guidelines for the pre saturated germicidal wipes including wearing gloves and other personal protective equipment when dispensing and using t...

Page 37: ...odule Place the end of the drain line into a container The water will passively drain into the container 2 Refill the reservoir Connect the fill tube From the Hypothermia therapy screen or the Normoth...

Page 38: ...e the Arctic Sun Temperature Management System requires periodic service and or replacement of key components The operator will be informed of preventative maintenance through the notification icon on...

Page 39: ...ronment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The Arctic Sun Temperature Management System uses RF energy only for its internal function Therefor...

Page 40: ...z 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE band 1 3 4 25 UMts Pulse mod 217 Hz 2 0 3 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE band 7 Pulse mod...

Page 41: ...to clear the line and then check to see if the flow has increased If faulty pad found replace faulty pad 5 To check whether a valve is leaking reinstall the pad into a different location on the fluid...

Page 42: ...that may affect cooling Are the machine vents blocked Is the room temperature too high Is the oxygen heated Are there hot lights or heating blankets on the patient Remove any impediments Monitor water...

Page 43: ...m is not working Call BD Customer Support If the water temperature increases to the commanded temperature the water heating system is working 5 Verify Patient Control System Performance Place system b...

Page 44: ...obe and temperature cable is secure 4 Check that the connection between the temperature cable and the Temp In 1connector is secure 5 Check the integrity of the temperature cable Flex the cable to chec...

Page 45: ...mental conditions that may affect cooling Are the machine vents blocked Is the room temperature too high Is the oxygen heated Are there hot lights or heating blankets on the patient Remove any impedim...

Page 46: ...g Call BD Customer Support If the water temperature increases to the commanded temperature the water heating system is working 5 Verify Patient Control System Performance Place system back into the pa...

Page 47: ...stem is unable to internally check the calibration of the Patient Temperature 2 channel within 1 0 C when the system is in Manual Control mode Patient Temperature 2 channel is made inactive 1 Power sy...

Page 48: ...riority Alarm One or more of the entries into the system event log is invalid The system event log is used by clinical engineering personnel for product service This issue does not affect the performa...

Page 49: ...ing is activated The system is not in Manual Control mode The high water temperature limit and low water temperature limits are set correctly See VI Operation Guide Therapy Settings High Water Limit o...

Page 50: ...water temperature remains below 10 C 50 F then the patient is generating excessive heat most likely from shivering which may or may not be visible Consider administration of additional medication for...

Page 51: ...and turn Control Module On 2 If alarm persists contact BD Customer Support 77 Non Recoverable System Error Medium Priority Alarm Chiller water temperature sensor out of range high resistance 1 Turn Co...

Page 52: ...has reached the end of its set duration and Rewarming Begins in Hypothermia Settings is set to Manually See VI Operation Guide Therapy Settings Rewarming Begins for more information Press the Start b...

Page 53: ...d period of time Prolonged warm water exposure may increase the risk for skin injury Assess patient s skin underneath the ArcticGel pads 1 Verify the custom parameters Patient target is set to the cor...

Page 54: ...erature by an independent measurement 3 Verify that the rate of temperature change is intentionally set to a low value 119 Patient Temperature 1 Outside of Monitor Mode Limits Low Priority Alarm Patie...

Page 55: ...S ARISING OUT OF HANDLING OR USE OF BARD EQUIPMENT OR DISPOSABLES EVEN IF BARD HAS BEEN ADVISED OF THE POSSIBILITY OR LIKELIHOOD OF SUCH DAMAGES IN NO EVENT WILL BARD S LIABILITY UNDER THIS WARRANTY E...

Page 56: ...Free 866 840 9776 Urgent Clinical Help arcticsun crbard com 2019 BD BD and the BD Logo are trademarks of Becton Dickinson and Company Bard ArcticGel Arctic Sun and Medivance are trademarks and or regi...

Page 57: ...The current Operator s Manual can be found at www medivance com manuals English 57 PK2800300 MM YYY...

Reviews: